Intrakranielle Gefäße: Arterien - Protokoll

Intrakranielle Gefäße: Arterien - Protokoll. 64x0,6mm Spiral-Scan von distaler ACC bis hoch parietal, Pitch 0,9, Rotationszeit: 0,33 s, Scanrichtung kaudokranial 120 kV, Eff. mAs 175 KM: 60 ml, 400 mg/ml, 4 ml/s; 40 ml NaCl 0,9% 4ml/s Bolustracking: ACC, Delay: 2 s

lavina
Download Presentation

Intrakranielle Gefäße: Arterien - Protokoll

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Intrakranielle Gefäße: Arterien - Protokoll • 64x0,6mm Spiral-Scan von distaler ACC bis hoch parietal, Pitch 0,9, Rotationszeit: 0,33 s, Scanrichtung kaudokranial • 120 kV, Eff. mAs 175 • KM: 60 ml, 400 mg/ml, 4 ml/s; 40 ml NaCl 0,9% 4ml/s • Bolustracking: ACC, Delay: 2 s • Rekon-Schichtdicke: 0,75 mm, Rekon-Inkrement: 0,5 mm, Rekon-Kernel: sehr weich, Matrix: 512 x 512 http://www.radiologie.ruhr-uni-bochum.de/

  2. Intrakranielle Gefäße: Arterien: Dual-Energy - Protokoll • 2x64x0,6mm Spiral-Scan von distaler ACC bis hoch parietal, Pitch 0,9, Rotationszeit: 0,33 s, Scanrichtung kaudokranial • 140 kV + 80 kV, Eff. mAs 60/255 • KM: 60 ml, 400 mg/ml, 4 ml/s; 40 ml NaCl 0,9% 4ml/s • Bolustracking: ACC, Delay: 2 s • Rekon-Schichtdicke: 0,75 mm, Rekon-Inkrement: 0,5 mm, Rekon-Kernel: sehr weich, Matrix: 512 x 512 http://www.radiologie.ruhr-uni-bochum.de/

  3. Intrakranielle Gefäße: Venen - Protokoll • 64x0,6mm Spiral-Scan von Atlasbogen bis hoch parietal, Pitch 0,9, Rotationszeit: 0,33 s, Scanrichtung kraniokaudal • 120 kV, Eff. mAs 175 • KM: 60 ml, 400 mg/ml, 4 ml/s; 40 ml NaCl 0,9% 4ml/s • Bolustracking: ACC, Delay: 10 s • Rekon-Schichtdicke: 0,75 mm, Rekon-Inkrement: 0,5 mm, Rekon-Kernel: sehr weich, Matrix: 512 x 512 http://www.radiologie.ruhr-uni-bochum.de/

  4. Halsgefäße Protokoll • 64x0,6mm Spiral-Scan von Aortenbogen bis Schädelbasis, Pitch 0,9, Rotationszeit: 0,33 s, Scanrichtung kaudokranial • 120 kV, Röhrenstrommodulation mit Ref.-mAs: 150 • KM: 60 ml, 400 mg/ml, 4 ml/s; 40 ml NaCl 0,9% 4ml/s • Bolustracking: Aortenbogen, Delay: 2 s • Rekon-Schichtdicke: 1,0 mm, Rekon-Inkrement: 0,7 mm, Rekon-Kernel: sehr weich, Matrix: 512 x 512 http://www.radiologie.ruhr-uni-bochum.de/

  5. Halsgefäße Dual-Energy - Protokoll • 2x64x0,6mm Spiral-Scan von Aortenbogen bis Schädelbasis, Pitch 0,9, Rotationszeit: 0,33 s, Scanrichtung kaudokranial • 140 kV + 80 kV, Röhrenstrommodulation mit Ref.-mAs: 70/300 • KM: 60 ml, 400 mg/ml, 4 ml/s; 40 ml NaCl 0,9% 4ml/s • Bolustracking: Aortenbogen, Delay: 2 s • Rekon-Schichtdicke: 1,0 mm, Rekon-Inkrement: 0,7 mm, Rekon-Kernel: sehr weich, Matrix: 512 x 512 http://www.radiologie.ruhr-uni-bochum.de/

  6. Thorax: Lungenarterien - Protokoll • 64x0,6mm Spiral-Scan von Jugulum bis Lungenende, Pitch 1,2, Rotationszeit: 0,33 s, Scanrichtung kraniokaudal • 120 kV, Röhrenstrommodulation mit Ref.-mAs: 150 • KM: 80 ml, 400 mg/ml, 4 ml/s; 40 ml NaCl 0,9% 4ml/s • Bolustracking: Truncus pulmonalis, Delay: 6 s • Rekon-Schichtdicke: 2,0 mm, Rekon-Inkrement: 1,5 mm, Rekon-Kernel: weich, Matrix: 512 x 512 http://www.radiologie.ruhr-uni-bochum.de/

  7. Thorax/Abdomen: Aorta - Protokoll • 64x0,6mm Spiral-Scan von Jugulum bis Beckenbodenende, Pitch 0,9, Rotationszeit: 0,33 s, Scanrichtung kraniokaudal • 120 kV, Röhrenstrommodulation mit Ref.-mAs: 120 • KM: 100 ml, 400 mg/ml, 4 ml/s; 40 ml NaCl 0,9% 4ml/s • Bolustracking: Aorta ascendens, Delay: 6 s • Rekon-Schichtdicke: 1,5 mm, Rekon-Inkrement: 1,0 mm, Rekon-Kernel: mittel weich, Matrix: 512 x 512 http://www.radiologie.ruhr-uni-bochum.de/

  8. Thorax: „Triple Rule-Out“ - Protokoll • 64x0,6mm Spiral-Scan von Jugulum bis Lungenende, Pitch 0,31, Rotationszeit: 0,33 s, Scanrichtung kraniokaudal, EKG-Gating • 120 kV, Röhrenstrommodulation mit Qual. Ref.-mAs/Rot.: 320 • KM: 70 ml, 400 mg/ml, 5 ml/s; 50 ml (30% KM 400mg/ml, 70% NaCl 0,9%), 5 ml/s; 50 ml NaCl 0,9%, 5 ml/s • Bolustracking: Aorta ascendens, Delay: 6 s • Rekon-Schichtdicke: 0,75 mm, Rekon-Inkrement: 0,5 mm, Rekon-Kernel: weich, Matrix: 512 x 512 http://www.radiologie.ruhr-uni-bochum.de/

  9. Abdomen: Nieren- und Mesenterialarterien - Protokoll • 64x0,6mm Spiral-Scan von Zwerchfell bis Beckenbodenende, Pitch 0,9, Rotationszeit: 0,33 s, Scanrichtung kraniokaudal • 120 kV, Röhrenstrommodulation mit Ref.-mAs: 120 • KM: 90 ml, 400 mg/ml, 4 ml/s; 40 ml NaCl 0,9% 4ml/s • Bolustracking: Aorta descendens, Delay: 6 s • Rekon-Schichtdicke: 0,75 mm, Rekon-Inkrement: 0,5 mm, Rekon-Kernel: mittel weich, Matrix: 512 x 512 http://www.radiologie.ruhr-uni-bochum.de/

  10. Extremitäten: Protokoll (Bauch-, Becken-, Beingefäße) • 64x0,6mm Spiral-Scan von Nierenoberpol bis OSG, Pitch 0,8, Rotationszeit: 0,33 s, Scanrichtung kraniokaudal • 120 kV, Röhrenstrommodulation mit Ref.-mAs: 120 • KM: 100 ml, 400 mg/ml, 4 ml/s; 60 ml NaCl 0,9% 4ml/s • Bolustracking: Aorta descendens, Delay: 2 s • Rekon-Schichtdicke: 1,5 mm, Rekon-Inkrement: 1,0 mm, Rekon-Kernel: mittel weich, Matrix: 512 x 512 http://www.radiologie.ruhr-uni-bochum.de/