Troubleshooting part i l.jpg
This presentation is the property of its rightful owner.
Sponsored Links
1 / 54

Troubleshooting Part I PowerPoint PPT Presentation


  • 514 Views
  • Uploaded on
  • Presentation posted in: General

Troubleshooting Part I. Objectives:. Understand the four basic steps used to solve troubleshooting problems Identify ECG abnormalities that result from pacing system malfunction and pseudomalfunction Recognize data and resources available to aid in troubleshooting pacing system anomalies

Download Presentation

Troubleshooting Part I

An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -

Presentation Transcript


Troubleshooting part i l.jpg

TroubleshootingPart I


Objectives l.jpg

Objectives:

  • Understand the four basic steps used to solve troubleshooting problems

  • Identify ECG abnormalities that result from pacing system malfunction and pseudomalfunction

  • Recognize data and resources available to aid in troubleshooting pacing system anomalies

  • Discern pacemaker functions that can affect patient hemodynamics

  • Describe the causes of pacemaker system anomalies and propose a potential solution


The steps used in troubleshooting are simple and remain the same for each type of problem l.jpg

The Steps Used in Troubleshooting Are Simple and Remain the Same for Each Type of Problem

  • Define the problem

  • Identify the cause of the problem

  • Correct the problem

  • Verify the solution


Defining the problem and identifying the cause l.jpg

Defining the Problem and Identifying the Cause


Potential problems identifiable on an ecg can generally be assigned to five categories l.jpg

Potential Problems Identifiable on an ECG Can Generally Be Assigned to Five Categories:

  • Undersensing

  • Oversensing

  • Noncapture

  • No output

  • Pseudomalfunction


Undersensing l.jpg

Undersensing

  • An intrinsic depolarization that is present, yet not seen or sensed by the pacemaker

P-wavenot sensed

Atrial Undersensing


Undersensing may be caused by l.jpg

Undersensing May Be Caused By:

  • Inappropriately programmed sensitivity

  • Lead dislodgment

  • Lead failure:

    • Insulation break; conductor fracture

  • Lead maturation

  • Change in the native signal


Oversensing l.jpg

Oversensing

  • The sensing of an inappropriate signal

    • Can be physiologic or nonphysiologic

...Though no activity is present

Marker channel shows intrinsic activity...

Ventricular Oversensing


Oversensing may be caused by l.jpg

Oversensing May Be Caused By:

  • Lead failure

  • Poor connection at connector block

  • Exposure to interference


Noncapture is exhibited by l.jpg

Noncapture is Exhibited By:

  • No evidence of depolarization after pacing artifact

Loss of capture


Noncapture may be caused by l.jpg

Noncapture May Be Caused By:

  • Lead dislodgment

  • Low output

  • Lead maturation

  • Poor connection at connector block

  • Lead failure


Less common causes of noncapture may include l.jpg

Less Common Causes of Noncapture May Include:

  • Twiddler’s syndrome

  • Electrolyte abnormalities – e.g., hyperkalemia

  • Myocardial infarction

  • Drug therapy

  • Battery depletion

  • Exit block


No output l.jpg

No Output

  • Pacemaker artifacts do not appear on the ECG; rate is less than the lower rate

Pacing output delivered; no evidence of pacing spike is seen


No output may be caused by l.jpg

No Output May Be Caused By:

  • Poor connection at connector block

  • Lead failure

  • Battery depletion

  • Circuit failure


Pseudomalfunctions l.jpg

Pseudomalfunctions

Pseudomalfunctions are defined as:

  • Unusual

  • Unexpected

  • Eccentric

    ECG findings that appear to result from pacemaker malfunction but that represent normal pacemaker function


Pseudomalfunctions may be classified under the following categories l.jpg

Pseudomalfunctions May Be Classified Under the Following Categories:

  • Rate

  • AV interval/refractory periods

  • Mode


Rate changes may occur due to normal device operation l.jpg

Rate Changes May Occur Due to Normal Device Operation:

  • Magnet operation

  • Timing variations

    • A-A versus V-V timing

  • Upper rate behavior

    • Pseudo-Wenckebach; 2:1 block

  • Electrical reset

  • Battery depletion

  • PMT intervention

  • Rate response


Magnet operation l.jpg

Magnet Operation

  • Magnet application causes asynchronous pacing at a designated “magnet” rate


A to a vs v to v timing l.jpg

A to A = 1000 ms

A to A = 1000 ms

AV = 150

AV = 200

V-A = 850

V-A = 800

A to A = 1000 ms

A to A = 950 ms

A to A vs. V to V Timing

A-A

Timing

Atrial rate is held constant at 60 ppm

AV = 150

AV = 200

V-V

Timing

V-A = 800

V-A = 800

Atrial rate varies with intrinsic ventricular conduction


Upper rate behavior l.jpg

Upper Rate Behavior

  • Pseudo-Wenckebach operation will cause a fluctuation in rate


Upper rate behavior21 l.jpg

Upper Rate Behavior

  • 2:1 block operation will cause a drastic drop in rate


Electrical reset and battery depletion l.jpg

Electrical Reset and Battery Depletion

  • Reset may occur due to exposure to electromagnetic interference (EMI) – e.g., electrocautery, defibrillation, causing reversion to a “back-up” mode

    • Rate and mode changes will occur

    • Device can usually be reprogrammed to former parameters

  • Elective replacement indicators (ERI) can resemble back-up mode

    • Interrogating device will indicate ERI (“Replace Pacer”)


Pmt intervention l.jpg

PMT Intervention

  • Designed to interrupt a Pacemaker-Mediated Tachycardia


Rate responsive pacing l.jpg

Rate Responsive Pacing

  • An accelerating or decelerating rate may be perceived as anomalous pacemaker behavior

VVIR / 60 / 120


Rate changes may occur due to therapy specific device operation l.jpg

Rate Changes May Occur Due to Therapy-Specific Device Operation

  • Hysteresis

  • Rate drop response

  • Mode switching

  • Sleep function


Hysteresis l.jpg

Hysteresis

  • Allows a lower rate between sensed events to occur; paced rate is higher

Hysteresis Rate 50 ppm

Lower Rate 70 ppm


Rate drop response l.jpg

Rate Drop Response

  • Delivers pacing at high rate when episodic drop in rate occurs

    • Pacing therapy indicated for patients with neurocardiogenic syncope


Mode switching l.jpg

Mode Switching

  • Device switches from tracking (DDDR) to nontracking (DDIR) mode


Sleep function l.jpg

Lower

Rate

Rate

Sleep

Rate

Sleep Function

30

mins.

30

mins.

Wake Time

Bed Time

Time


Av intervals refractory periods may appear anomalous due to l.jpg

AV Intervals/Refractory Periods May Appear Anomalous Due to:

  • Safety pacing

  • Blanking

  • Rate-adaptive AV delay

  • Sensor-varied PVARP

  • PVC response

  • Noncompetitive atrial pace (NCAP)


Safety pacing l.jpg

Ventricular Safety Pace

Safety Pacing

  • Designed to prevent inhibition due to “crosstalk”

    • Delivers a ventricular pace 110 ms after an atrial paced event


Blanking l.jpg

Blanking

DDDR / 60 / 125 / 200 / 225


Rate adaptive av delay l.jpg

Rate-Adaptive AV Delay

  • AV interval shortens as rate increases

PAV delay with no activity: 150 ms

PAV with activity: 120 ms


Sensor varied pvarp l.jpg

Sensor-Varied PVARP

  • PVARP will shorten as rate increases

Long PVARP with little activity

Shorter PVARP with increased activity


Pvc response l.jpg

PVC Response

  • PVARP will extend to 400 ms

DDD / 60 / 120 PVARP 310 ms


Noncompetitive atrial pace ncap l.jpg

Noncompetitive Atrial Pace (NCAP)

  • Prevents atrial pacing from occurring too close to relative refractory period, which may trigger atrial arrhythmias


A change in pacing modes may be caused by l.jpg

A Change in Pacing Modes May Be Caused By:

  • Battery depletion indicators (ERI/EOL)

  • Electrical reset

  • Mode switching

  • Noise reversion


Noise reversion l.jpg

Noise Reversion

  • Sensing occurring during atrial or ventricular refractory periods will restart the refractory period. Continuous refractory sensing is called noise reversion and will:

    • Cause pacing to occur at the sensor-indicated rate for rate-responsive modes

    • Cause pacing to occur at the lower rate for non- rate-responsive modes


Noise reversion39 l.jpg

Noise Reversion


Slide40 l.jpg

Note: Adverse patient symptoms may occur as a result of any of the previously mentioned pacing system malfunctions and some pseudomalfunctions.


Management of patient symptoms may be necessary as a result of l.jpg

Management of Patient Symptoms May Be Necessary as a Result of:

  • Muscle stimulation

  • Palpitations

  • Pacemaker syndrome

  • Shortness of breath due to inappropriate rate response settings


Muscle stimulation may be caused by l.jpg

Muscle Stimulation May Be Caused By:

  • Inappropriate electrode placement near diaphragm or nerve plexus

  • Break in lead insulation

  • Unipolar pacing


Palpitations may manifest from l.jpg

Palpitations May Manifest From:

  • Pacemaker syndrome

  • Pacemaker-Mediated Tachycardia (PMT)


Pacemaker syndrome l.jpg

Pacemaker Syndrome

“An assortment of symptoms related to the adverse hemodynamic impact from the loss of AV synchrony.”


Pacemaker syndrome45 l.jpg

Pacemaker Syndrome

Symptoms include:

  • Dizziness

  • Presyncope

  • Chest tightness

  • Shortness of breath

  • Neck pulsations

  • Apprehension/malaise

  • Fatigue


Pacemaker syndrome may be caused by l.jpg

Pacemaker Syndrome May Be Caused By:

  • Loss of capture, sensing

  • A-V intervals of long duration

  • Onset of 2:1 block

  • Single chamber system

  • Absence of rate increase with exercise


Pacemaker mediated tachycardia pmt l.jpg

Pacemaker-Mediated Tachycardia (PMT)

  • A rapid paced rhythm that can occur with atrial tracking pacemakers


Pmt is the result of l.jpg

PMT is the Result of:

  • Retrograde conduction

  • Tracking fast atrial rates (physiologic or non-physiologic)


Retrograde conduction l.jpg

Retrograde Conduction


Retrograde conduction may be caused by l.jpg

Retrograde Conduction May Be Caused By:

  • Loss of A-V synchrony due to:

    • Loss of sensing/capture

    • Myopotential sensing

    • Premature ventricular contraction (PVC)

    • Magnet application


High rate atrial tracking is caused by l.jpg

High Rate Atrial Tracking is Caused By:

  • Supra-ventricular tachyarrhythmias

  • Atrial oversensing


Slide52 l.jpg

General Medtronic Pacemaker Disclaimer

INDICATIONS

Medtronic pacemakers are indicated for rate adaptive pacing in patients who may benefit from increased pacing rates concurrent with increases in activity (Thera, Thera-i, Prodigy, Preva and Medtronic.Kappa 700 Series) or increases in activity and/or minute ventilation (Medtronic.Kappa 400 Series).

Medtronic pacemakers are also indicated for dual chamber and atrial tracking modes in patients who may benefit from maintenance of AV synchrony. Dual chamber modes are specifically indicated for treatment of conduction disorders that require restoration of both rate and AV synchrony, which include various degrees of AV block to maintain the atrial contribution to cardiac output and VVI intolerance (e.g., pacemaker syndrome) in the presence of persistent sinus rhythm.

9790 Programmer

The Medtronic 9790 Programmers are portable, microprocessor based instruments used to program Medtronic implantable devices.

9462

The Model 9462 Remote Assistant™ is intended for use in combination with a Medtronic implantable pacemaker with Remote Assistant diagnostic capabilities.

CONTRAINDICATIONS

Medtronic pacemakers are contraindicated for the following applications:

·       Dual chamber atrial pacing in patients with chronic refractory atrial tachyarrhythmias.

·       Asynchronous pacing in the presence (or likelihood) of competitive paced and intrinsic rhythms.

·       Unipolar pacing for patients with an implanted cardioverter-defibrillator because it may cause unwanted delivery or inhibition of ICD therapy.

·       Medtronic.Kappa 400 Series pacemakers are contraindicated for use with epicardial leads and with abdominal implantation.

WARNINGS/PRECAUTIONS

Pacemaker patients should avoid sources of magnetic resonance imaging, diathermy, high sources of radiation, electrosurgical cautery, external defibrillation, lithotripsy, and radiofrequency ablation to avoid electrical reset of the device, inappropriate sensing and/or therapy.

9462

Operation of the Model 9462 Remote Assistant™ Cardiac Monitor near sources of electromagnetic interference, such as cellular phones, computer monitors, etc. may adversely affect the performance of this device.

See the appropriate technical manual for detailed information regarding indications, contraindications, warnings, and precautions.

 Caution: Federal law (U.S.A.) restricts this device to sale by or on the order of a physician.


Slide53 l.jpg

Medtronic Leads

For Indications, Contraindications, Warnings, and Precautions for Medtronic Leads, please refer to the appropriate Leads Technical Manual or call your local Medtronic Representative.

Caution: Federal law restricts this device to sale by or on the order of a Physician.

Note:

This presentation is provided for general educational purposes only and should not be considered the exclusive source for this type of information. At all times, it is the professional responsibility of the practitioner to exercise independent clinical judgment in a particular situation.


Continued in troubleshooting part ii l.jpg

Continued inTroubleshootingPart II


  • Login