Microscopy overview of different methods
Download
1 / 59

Microscopy: Overview of Different Methods - PowerPoint PPT Presentation


  • 290 Views
  • Updated On :

Microscopy: Overview of Different Methods. EML 5930 (27-750) Advanced Characterization and Microstructural Analysis A. D. Rollett, P.N Kalu Spring 2008. (MACRO)TEXTURE. Review : Recall that X-ray texture (macrotexture):

Related searches for Microscopy: Overview of Different Methods

loader
I am the owner, or an agent authorized to act on behalf of the owner, of the copyrighted work described.
capcha
Download Presentation

PowerPoint Slideshow about 'Microscopy: Overview of Different Methods' - julie


An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -
Presentation Transcript
Microscopy overview of different methods

Microscopy: Overview of Different Methods

EML 5930 (27-750)

Advanced Characterization and Microstructural Analysis

A. D. Rollett, P.N Kalu

Spring 2008


Macro texture
(MACRO)TEXTURE

Review:

Recall that X-ray texture (macrotexture):

  • Provides an overview of the crystallographic texture of material - only texture information is obtained.

  • Provides the volume fraction of the specimen, which has a particular orientation.

  • Does not tell how grains are distributed in the material.



Microtexture
MICROTEXTURE microstructure.

  • Microtexture technique provides concurrently the spatial location and the orientation of individual grains in a sample.

  • This technique can be referred to as the modern approach to texture investigation.

  • Until recently, the orientation of individual crystals can only be determined by Selected Area Diffraction Pattern (SADP) technique on TEM - very tedious.


  • Phenomena that can be investigated using microtexture: microstructure.

    • Effect of property variation as a function of orientation

    • Misorientation between neighboring grains or distribution of grain boundary geometry, which can result in the grain boundary/property relationship.

    • Correlations between geometrical and orientational parameters of the grains

    • Orientation variations within individual grains

    • Phase Identification/Relationships

    • Direct ODF measurement



Diffraction: diffraction of radiation by a crystal lattice.

  • Crystal structure analysis is usually based on diffractionphenomena caused by the interaction of matter with X-rays, electrons, or neutrons.

  • Therefore, when either X-rays or electrons interact with crystalline material, they are:

    (a) Subject to diffraction – have similar wave properties.

    (b) Monochromatic radiation - produce a series of strongly diffracted beams leaving the crystal in defined and predicted directions.


  • The resultant diffraction pattern is given by Braggs Law, and this is given by …….. 1

    where, d: Interplanar spacing

    : Grazing angle of incidence (Bragg angle)

    n: Integer (0, 1, 2, 3 ….. )

    : Wavelength of the incident electrons

  • Note

    • With diffraction, we use Reciprocal lattice in which sets of lattice planes are represented simply by a set of points in reciprocal space.


The Reciprocal Lattice and this is given by …….. 1

  • Very useful in metric calculations

  • Let a, b, c be the elementary translations of a space lattice (direct lattice)

  • A second lattice, reciprocal to the first one, is defined by translations a*, b*, c*, which satisfy the following conditions:

    a*. b = a*. c = b*. a = b*. c = c*. a = c*. b = 0

    and

    a*. a = b*. b = c*. c = 1

  • Equation 2a suggests that a*is normal to the plane (b,c),

    b*is normal to the plane (a,c), and

    c*is normal to the plane (a,b),

(2a)

(2b)


  • The magnitude and sense of and this is given by …….. 1a*, b*, c*are fixed by (2b)

  • According to (2a), a* may be written as:

    a* = p (b ^ c)

    where p is a constant.

    The value of p is obtained by if the scalar product of both sides of (3a) by a is taken:

    a* . a = 1 = p (b ^ c . a) = pV

  • Therefore, p = 1/V, and equation (3a) can be written as:

    a* = { (b ^ c)}1/V ………

(3a)

(3b)


Table 1. and this is given by …….. 1The Characteristics of Light and Various Radiations Used for Texture Measurement by Diffraction.

  • Electrons are the only radiation in which their penetration depth and interaction volume is small enough to allow diffraction from individual grains (very small volume).

  • Therefore, only electrons can be used for MICROTEXTURE


  • Microtexture technique can either be TEM-based or SEM-based. and this is given by …….. 1

  • Until mid-1980, the TEM-based was the major microtexture technique, although an SEM-based Selected Area Channeling technique was available.

  • Modern SEM-based technique known as can now be classified into two:

    • manual

    • automated


Principles of electron microscopy
Principles of Electron Microscopy and this is given by …….. 1


Microscopy
MICROSCOPY and this is given by …….. 1

Introduction:

  • Several new microscopy-based microcharacterization techniques have been developed over the last four decades, which have significantly extended the ability to study the microstructure of materials.

  • In addition to Optical Metallography, there is a range of Electron Optical techniques.

  • Electron microscope (developed in 1931) was initially used for the study of biological systems, but thin foil techniques in the mid-1950’s enabled microstructural investigations to be undertaken on metals and alloys.


Typical Information from Electron Microscope: and this is given by …….. 1

  • Chemical composition of materials can be obtained using electron microprobes to produce characteristic X-ray emissions and electron energy losses.

  • Imaging (surface) can be characterized using secondary electrons, backscattered electrons, photo-electron, Auger electrons and ion scattering.

  • Crystallography or crystal structure information can be obtained from backscattered electrons (diffraction of photons or electrons).

    The various studies of materials exploit at least one of the above information, as well as the excellent spatial resolution of electron microscopes.


Figure 1 and this is given by …….. 1. Summary of the various signals obtained by interaction of electrons with matter in an electron microscope


Comparison between optical and electron microscopy
Comparison between Optical and Electron Microscopy: and this is given by …….. 1

  • In many ways, electron microscopes (Scanning and Transmission) are analogous to light microscopes.

  • Fundamentally and functionally, electron microscopes (EM) and optical microscopes (OM) are identical.

  • That is, both types of microscopes serve to magnify minute objects normally invisible to the naked eyes.

  • Basically, component terminology of an EM is similar to that of an OM. Both microscopes consist the following (see Figures 2 and 3):


Figure 2. and this is given by …….. 1A simple optical, transmission microscope system comprising a condenser and objective lens.


Figure 3. and this is given by …….. 1Comparison of image formation.


(a) and this is given by …….. 1Source of Illumination as light source

  • Electron Gun produces an electron beam by thermionic or field emission - EM

  • Lamp produces light beam (including uv rays) - OM

    (b) Condenser Lens system projects a near parallel radiation on to the specimen

  • Electro-magnets of variable focal length are the lenses in EM.

  • Curved transparent substance - OM

    (c) Series of Imaging Lenses form the Image of the specimen



Electron microscopes
ELECTRON MICROSCOPES between the two types of microscopes, a detailed comparison is provided in Table 1.


ELECTRON EMISSION between the two types of microscopes, a detailed comparison is provided in Table 1.

  • The liberation of electrons from the surface of a solid into vacuum.

  • The process of electron emission is similar to that of ionization of a free atom

    • The energy of the electrons in an atom is lower than that of an electron at rest in vacuum; consequently, in order to ionize an atom, energy must be supplied to the electrons in some way or other.

  • A solid can only emit electrons if some of the electrons have energies equal to, or larger than, that of an electron at rest in vacuum. This may be achieved by various means, such as by

    • heating, irradiation with light (photoemission),

    • bombardment with charged particles (secondary emission), or

    • using of a strong electric field (field, or cold, emission).


ELECTRON SOURCES between the two types of microscopes, a detailed comparison is provided in Table 1.

  • Electron sources in electron beam instruments are required to provide either

    • a large total current beam of about 50 m diameter - low magnification TEM, or

    • a high intensity probe of electrons as small as 0.5 nm in diameter - SEM

  • There are three different types of electron source available


a) between the two types of microscopes, a detailed comparison is provided in Table 1.Thermionic tungsten hairpin filament

This is usually heated to about 2800 K by direct resistance heating. The surrounding grid, known as the Wehnelt cylinder, and the anode, which is at earth potential, act as an electrostatic lens.

For an operating condition of 100 kV, the brightness is about

3 x 105 A cm-2 sr-1.


(a) between the two types of microscopes, a detailed comparison is provided in Table 1.


(b) between the two types of microscopes, a detailed comparison is provided in Table 1.

Figure 4. Schematic diagram of a conventional tungsten thermionic source. (a) the filament F and Wehnelt cylinder (b) schematic ray path showing focusing action.


b) between the two types of microscopes, a detailed comparison is provided in Table 1.Lanthanum hexaboride crystal (LaB6)

The only difference between the conventional assembly illustrated in Figure 4 and a modern LaB6 assembly is that extra pumping holes are present in the Wehnelt cap to ensure a better pumping speed near the tip.

Higher current (greater than the tungsten) is obtainable in small probes.

The brightness of a LaB6 can be as high as 107 A cm-2 sr-1 at 100kV.


c) between the two types of microscopes, a detailed comparison is provided in Table 1.Field emission source

  • The emission of electrons from a metal or semiconductor into vacuum under the influence of a strong electric field.

  • Field emission - electrons tunnel through a potential barrier, rather than escaping over it as in thermonic emission.

    • The effect is purely quantum-mechanical, with no classical analog.


  • In most cases, it is a <111> orientation crystal of tungsten, and a Wehnelt cylinder, which is raised to an extraction potential up to about 4 kV in order to cause emission from the tip of the crystal.

  • There is a requirement of high vacuum for this source

  • The brightness of cold or thermal emission source can be about 104 times of a conventional tungsten filament.

  • The high brightness of this source make them preferable for scanning instruments.

  • Cold, thermal or Schottky


Schotty field emission source electron gun for sem
Schotty Field Emission Source Electron Gun for SEM tungsten, and a Wehnelt cylinder, which is raised to an extraction potential up to about 4 kV in order to cause emission from the tip of the crystal.

  • Source:

    • Schotty field emitter (ZrO/W

    • High brightness

    • Highly-stable electron beam

  • High current density

  • 100 nm spot size at 5 nA sample current

  • The field emission source – current density can be maximize and still produce a 100 nm spot size at 5 nA.

    • Optimised for Auger analysis at very high spatial resolution with SEM.


Electron lenses and optics
ELECTRON LENSES AND OPTICS tungsten, and a Wehnelt cylinder, which is raised to an extraction potential up to about 4 kV in order to cause emission from the tip of the crystal.


  • Electron lenses in microscopes are generally electromagnetic.

  • There are three types of magnetic lenses in uses (refer to Figure 5):

    (a) a multi-layer coil; i.e., an air-core solenoid coil

    (b) a coil enclosed by soft iron plates (in order to reduce leakage flux) containing a gap (in order to concentrate the induction field) and

    (c) a coil enclosed by soft iron plates containing a gap and internal soft iron pole pieces (in order to ensure a high intensity magnetic field)


Figure 5. electromagnetic. Types of magnetic electron lenses.


  • Almost all modern electron microscopes use pole pieces for high resolving power and high magnification.

  • The function of such an electron lens is more or less the same as that of horse-shoe magnets symmetrically arranged about an axis.

  • Accordingly, all the parallel electron beams incident to the curved magnetic field converge at one point.


Electron optics
ELECTRON OPTICS high resolving power and high magnification.

  • The action of a magnetic field on an electron is described by a well-known right hand rule where thumb, first and second finger are used to represent the terms in a vector product

  • The force F which an electron of charge -e experiences when travelling with velocity v, due to a magnetic field B, is given by:

    …………(4a)


Figure 6 high resolving power and high magnification. . Schematic diagram of the action of a cylindrical magnetic lens on the path of non-axial electrons.


and the magnitude of the force is then given by high resolving power and high magnification.

…………..(4b)

where  is the angle between B and v.

If the initial velocity of an electron is divided into two components, vp parallel to B and vo perpendicular to B, then the value of vp will be unchanged by B (since  will be zero) and the force resulting from B and vo will result in a circular motion of the electron about B (see Figures 7 and 8).


Figure 7. high resolving power and high magnification. Electrons passing through magnetic lens


  • Figs. 7 & 8 shows the trajectory of an electron passing through such a magnetic field.

  • Although the electron beam path in a magnetic lens is not the same as the light ray path in an optical lens, the results are similar.

  • As shown in Figs. 7 & 8, the electron travels rectilinearly, crosses the axis, moves through the magnetic field along a spiral orbit, approaches the axis, crosses the axis again, and travels rectilinearly.


Figure 8 through such a magnetic field.. Schematic diagram showing the trajectory of an electron through a magnetic lens.


The radius r of this spiral motion is given by: through such a magnetic field.

……….(4c)

Since generally, in electron microscopes, electron beams near the axis are used for forming an image, a is extremely small.

This effect is similar to the converging action of an optical convex lens, and if the revolution of the electron about the axis is omitted, the converging action of an electron lens can be considered to be identical to that of an optical lens.


  • A magnetic lens containing pole pieces magnetized to near-saturation for concentrating magnetic flux in a very narrow space constitutes a thin lens.

  • The focal length f and rotation angle q are given as follows:

    …….. 5

    where, V => Accelerating voltage

     = specific charge of electron e/m


RESOLUTION OF LENS near-saturation for concentrating magnetic flux in a very narrow space constitutes a thin lens.

  • Resolution defines the smallest separation of two points in the object, which may be distinctly reproduced in the image.

    Light Microscope: The resolving power for light microscope is determined by diffraction aberration and can be defined as 1,2

    ……….. 6

  • where  is the wavelength of the illumination,

    n is the refractive index of the medium in front of the lens,

     is the semi-angle (aperture angle) subtended by the object at the lens

    NA = n sinα= numerical aperture = measure of light-gathering ability

    = 0.95 (max. with air). Higher (~1.515) for oil-immersion objtvs.

    k is a constant usually taken to be 0.61.


Optical Microscope near-saturation for concentrating magnetic flux in a very narrow space constitutes a thin lens. –

 = 50 nm (for white light Illumination)

n sin  = 0.135 (when fitted with an oil immersion lens)

Therefore, it is possible to achieve a resolution of about 250 nm in Optical Microscopes.

  • Filters can also be used to enhance the resolving power of an objective. For light:

    • The shorter wavelengths are at the violet-blue-green end of the spectrum

    • The higher wavelengths are at the orange-red of the spectrum.


Electron Microscope near-saturation for concentrating magnetic flux in a very narrow space constitutes a thin lens. –

  • De Broglie relationship relates the wavelength of electrons, , to their momentum, mv (m is the mass and v is velocity), by h – Planck’s constant, such that:

    ……… 7

  • Since electrons are accelerated by a potential difference of V volts and have a kinetic energy K, where K is given as –

    ……... 8

    Therefore,

    ……… 9


By considering equations (7) and (9), we have – near-saturation for concentrating magnetic flux in a very narrow space constitutes a thin lens.

……. 10

The energy term eV is expressed in electron volts and represents that energy required to pass an electron through a potential difference of one volt ( 1 eV = 1.602 x 10-19 J).


When the velocity of the electron approaches the speed of light, v c, a relativistic correction is required for the voltage, such that

………………….. 11

where mo is the mass of the electron at rest. It is important to use this correction for cases when V  105 volts. Table 2 presents a chart of electron wavelength in relation to applied voltage.


Table 2. light, v c, a relativistic correction is required for the voltage, such thatVariation of Electron Wavelength with applied voltage


  • If light, v c, a relativistic correction is required for the voltage, such thatdiffraction aberration is considered (see equation (6)), the resolving power of 100 and 300 keV electron microscopes will be about 0.0025 nm and 0.0017 nm respectively. These values are not realistic!

  • The ultimate resolution of an electron microscope is dictated by the defects in the imaging system rather than by the wavelength (diffraction aberration) of the radiation employed.


Lens defects: light, v c, a relativistic correction is required for the voltage, such that

  • Chromatic aberration is related to energy losses within the specimen. This is generally of little importance except for very thick regions of thin foils or bulk specimens.

  • Astigmatism may arise from

    • intrinsic defects in the objective lens, or

    • from contamination on the lens and the objective aperture.

      Microscope manufacturers have developed methods for correcting this defect

    • small set of coils are used to produce uniform magnetic field.


  • Spherical aberration light, v c, a relativistic correction is required for the voltage, such that is the main factor, which limits the performance of electromagnetic lenses used in microscopes.

    • It is a function of the lens design and the acceptance angle () for electrons entering the lens.

    • This angle must be kept at a minimum, however this decreases the resolution limit due to diffraction. The bestresolution is accomplished with a compromise value of , and is given by

      ……… 12

    • where cs is the coefficient of spherical aberration, and min is typically less than 3 Å on modern microscopes.


Current state-of-the-art TEM light, v c, a relativistic correction is required for the voltage, such that

capability

  • Analytical Probe sizes less than 1A

  • TEM resolution 0.8A or better

  • Holographic reconstructions less than 1A

  • FEG: cold/thermal/Schottky

  • Energy resolution in EELS less than 20meV – more typical values less than 100meV – routine values of 700 – 800meV. Monochromators help.

  • WDS resolution less than 1 eV (1000meV)

  • CCD Camera images 4K x 4K (8K under development at UCSD-commercial version just delivered to Fujiyoshi)

  • Contemporary computer control – contemporary means TODAY


Current state-of-the-art TEM light, v c, a relativistic correction is required for the voltage, such that

capability

  • Accelerating voltages up to 3.0 MeV

  • A variety of energy filters: Omega, Mandoline, Gamma, single-sector (GATAN)

  • Goniometer Automation (searching, recording, tilting (tomography/kspace tilting and navigation))

  • Detector Automation - autoinsertion, autoconfiguration of instrument modes

  • Piezo elements for nanopositioning

  • SPM – Nanofactory

  • Remote control – taking the operator away from the microscope


Resolution/Cs Correction light, v c, a relativistic correction is required for the voltage, such that

  • Before the Cs corrector became a practical invention, the TEM/STEM went through the final transition to an FEG instrument capable of uncorrected extended information limit and smaller, higher current probes.

  • This happened in the early part of 1990 and set the stage for the development of the TEM/STEM Cs corrector.


Directions for the future
Directions for the future: light, v c, a relativistic correction is required for the voltage, such that

  • Areas of Future directions for Improvements include:

  • resolution (image/analytical – spatial/energy), computer control, automation, vacuum, in-situ capability, remote control

  • The question is “How will these improvements be effected?”

  • Most instruments today are available with a variety of resolution enhancers:

    • Cs correctors for TEM and STEM image and analytical resolution

    • Monochromators for improved energy resolution.


Directions for the future1
Directions for the future: light, v c, a relativistic correction is required for the voltage, such that

  • Other subjects to consider include:

  • Tomography: Image series acquisition, Image reconstruction, and 3D rendering.

  • Lorentz: objective minilens with a resolution of about 2nm that can be used as an objective lens

  • Remote control

  • Data management

  • EDS and EELS/Imaging, and

  • Integration of all functions.


Today Electron Microscopes represent state-of-the-art technology incorporating:

  • Clean High vacuum and Ultra-high vacuum

  • Extreme electronic stability - lenses and high voltage

  • Machining accuracy to 1um or better, and

  • Unsurpassed mechanical stability

They remain the highest lateral spatial resolution spectroscopy devices in the world

They can currently resolve sub-angstrom features and can perform spectroscopy on single atomic columns or even single atoms.