Pacemaker troubleshooting single chamber pacemakers
Download
1 / 63

Pacemaker troubleshooting-single chamber pacemakers - PowerPoint PPT Presentation


  • 228 Views
  • Uploaded on

Pacemaker troubleshooting-single chamber pacemakers. Reasons for evaluation. Patient symptomatic Palpitation Syncope presyncope Pacemaker malfunction suspected ECG Telemetry Ambulatory ECG Routine pacemaker follow up. Patient details. Indication for pacing Implant operative note

loader
I am the owner, or an agent authorized to act on behalf of the owner, of the copyrighted work described.
capcha
Download Presentation

PowerPoint Slideshow about ' Pacemaker troubleshooting-single chamber pacemakers' - joseph-benjamin


An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -
Presentation Transcript

Reasons for evaluation
Reasons for evaluation

  • Patient symptomatic

    • Palpitation

    • Syncope

    • presyncope

  • Pacemaker malfunction suspected

    • ECG

    • Telemetry

    • Ambulatory ECG

  • Routine pacemaker follow up


Patient details
Patient details

  • Indication for pacing

  • Implant operative note

  • Diagnosis

  • Medication –can alter pacing tresholds

  • DC ,MRI,electrocautery

  • Trauma,electrical current exposure

  • Concurrent medical problems-CRF,hyperkalemia

  • Prior ECG,CXR


Pacemaker system

  • manufacturer

  • Model

  • current programming

  • date of implant

  • special features

  • Sensing and pacing tresholds

  • impedance


Lead system

  • manufacturer

  • model

  • polarity

  • Fixation

  • insulation and date


Causes of pacemaker malfunction
Causes of pacemaker malfunction

  • Pacing stimuli present with failure to capture

  • Pacing stimuli present with failure to sense

  • Pacing stimuli absent

    • Oversensing

    • No output

  • Pseudomalfunction


Failure to capture
Failure to capture

  • No evidence of depolarization after pacing artifact

Loss of capture


Early post device implantation
Early post device implantation

  • CXR abnormal

    • Lead dislodgement

      • Downward migration of PG

      • Failure to secure anchoring sleeves properly

      • Too little or too much slack

    • Lead perforation

  • Elevated impedance

    • Loose set screw

    • failure to seat lead pin properly in header

  • Normal CXR,normal impedance

    • Micro lead dislodgement

    • Inflammatory response


Late post device implantation
Late post device implantation

  • Battery depletion

  • Twiddler s syndrome

  • Abnormal myocardium

  • Insulation failure

  • Conductor failure

    • Mechanical stress on lead

      • Anchoring sleeve

      • Interaction with generator in pocket

      • b/w clavicle and first rib


  • Increase the energy in the output pulse

    • Run a capture threshold test

    • Adjust the output parameters, if necessary

      • Pulse amplitude (V)

      • Pulse width or duration (ms)

    • It is generally more efficient to increase the pulse amplitude

  • Investigate possible lead problems

  • Reprogram device polarity


Undersensing
Undersensing

  • An intrinsic depolarization that is present, yet not seen or sensed by the pacemaker

P-wavenot sensed

Atrial Undersensing


  • Undersensing occurs when the pacemaker does not detect intrinsic activity that really is there

  • Undersensing causes the pacemaker to pace more than it should


Undersensing may be caused by
Undersensing May Be Caused By:

  • Inappropriately programmed sensitivity

  • Lead dislodgment

  • Lead failure:

    • Insulation break; conductor fracture

  • Lead maturation

  • Change in the native signal

  • Functional

    • Magnet

    • Noise reversion


  • Adjust the sensitivity setting

    • Run a sensing threshold test

    • Measure the intrinsic signals

    • Adjust the sensitivity appropriately

      • To increase sensitivity, decrease the mV setting

    • Make all changes to sensitivity settings in small steps since large changes may only introduce new sensing problems


Oversensing
Oversensing

  • The sensing of an inappropriate signal

    • Can be physiologic or nonphysiologic

...Though no activity is present

Marker channel shows intrinsic activity...

Ventricular Oversensing


  • Oversensing occurs when the pacemaker inappropriately “thinks” that it sees intrinsic activity that is not there

  • Oversensing causes the pacemaker to inhibit the pacing output pulse, even though the device should be pacing


  • Myopotentials

    • Pectoralis,rectusabdominis,diaphragm

    • Bipolar leads are less susceptible

  • Oversensing of normal intracardiac signals

    • Reduce sensitivity of affected lead to eliminate oversensing

  • EMI

  • Provocative maneuvers may be needed if intermittent symptoms

  • Magnet eliminates pauses-oversensing


  • Reprogram the sensitivity

    • Conduct a sensing threshold test

    • Adjust the sensitivity by making the device less sensitive (increase the mV setting)

    • Make only small changes

  • Extend the refractory period


No output
No Output

  • Pacemaker artifacts do not appear on the ECG; rate is less than the lower rate

Pacing output delivered; no evidence of pacing spike is seen


No output may be caused by
No Output May Be Caused By:

  • Poor connection at connector block

  • Lead failure

  • Battery depletion

  • Circuit failure


  • Steps to take for possible loss of output

    • Verify all lead connections

    • Check lead integrity

    • Evaluate battery status

    • Contact the device manufacturer

  • Loss of output may require the replacement of all or part of the pacing system


Pseudomalfunctions
Pseudomalfunctions

Pseudomalfunctions are defined as:

Unusual,UnexpectedECG findings that appearto result from pacemaker malfunction but that represent normal pacemaker function


  • Hysteresis

  • Magnet rate

  • rate responsive pacing

  • Noise reversion

  • Rate drop response

  • Sleep rate algorithm


Hysteresis
Hysteresis

  • Allows a lower rate between sensed events to occur; paced rate is higher

Hysteresis Rate 50 ppm

Lower Rate 70 ppm


Magnet operation
Magnet Operation

  • Magnet application causes asynchronous pacing at a designated “magnet” rate


  • Threshold Margin Test (TMT)

    • Three beats at 100 bpm, followed by a magnet rate of 85

    • Third beat has an automatic pulse width decrement of 25%

    • Elective replacement indicators-change the rate from 85 to 65

  • Extended TMT.

    • TMT is performed at 100 ppm

    • Pulse width reduced by 25% on 3rd , 50% on 5th , and 75% on 7th


  • Rate responsive pacing
    Rate Responsive Pacing

    • An accelerating or decelerating rate may be perceived as anomalous pacemaker behavior

    VVIR / 60 / 120


    Electrical reset and battery depletion
    Electrical Reset and Battery Depletion

    • Reset may occur due to exposure to EMI

      electrocautery, defibrillation, causing reversion to a “back-up” mode

      • Rate and mode changes will occur

      • Device can usually be reprogrammed to former parameters

    • Elective replacement indicators (ERI) can resemble back-up mode

      • Interrogating device will indicate ERI (“Replace Pacer”)


    A change in pacing modes may be caused by
    A Change in Pacing Modes May Be Caused By:

    • Battery depletion indicators (ERI/EOL)

    • Electrical reset

    • Mode switching

    • Noise reversion


    Noise reversion
    Noise Reversion

    • Sensing occurring during atrial or ventricular refractory periods will restart the refractory period. Continuous refractory sensing is called noise reversion and will:

      • Cause pacing to occur at the sensor-indicated rate for rate-responsive modes

      • Cause pacing to occur at the lower rate for non- rate-responsive modes


    Noise reversion1
    Noise reversion

    VT in a patient with VVI-pacing occurs at lower rate due to noise reversion


    Rate drop response
    Rate drop response

    Delivers pacing at high rate when episodic drop in rate occurs


    Muscle stimulation may be caused by
    Muscle Stimulation May Be Caused By:

    • Inappropriate electrode placement near diaphragm or nerve plexus

    • Break in lead insulation

    • Unipolar pacing


    Flouro diagnosis
    Flouro diagnosis


    1.ECG shows

    A.Failure to sense

    B.Failure to capture

    C.Hysteresis

    D.Oversensing


    2.ECG shows

    A.Failure to sense

    B.Failure to capture

    C.Hysteresis

    D.Oversensing


    3.ECG shows

    A.Failure to sense

    B.Failure to capture

    C.Hysteresis

    D.Oversensing


    • 4.ECG demonstrates

      a.Failure to sense

      b.Failure to capture

      c.Functional non capture

      d.fusion


    5

    ECG shows A.Failureto sense

    B.Failureto capture

    C.Hysteresis

    D.Functionalnon capture


    6

    ECG shows

    A.Functional non capture

    B.Failure to capture

    C.Hysteresis

    D.Oversensing


    7.possibilities are

    • A.VVI in VOO mode

    • B.magnet kept

    • C.persistent ventricular undersensing

    • D.Noise reversion





    11.What are appropriate in this patient caused by

    A.sensitivity setting to be adjusted

    B.lower the programmed rate

    C.introduce hysteresis

    D.check capture treshold


    12.What is the next step in this patient caused by

    • A.adjust sensitivity-increase mV setting

    • B.adjust sensitivity-decrease mV setting

    • C.pacingtreshold test-adjust output parameters

    • D.activate hysteresis


    13.What caused byis appropriate in this patient

    • A.adjust sensitivity-increase mV setting

    • B.adjust sensitivity-decrease mV setting

    • C.pacingtreshold test-adjust output parameters

    • D.activate hysteresis


    14.What caused byis appropriate in this patient

    • A.adjust sensitivity-increase mV setting

    • B.adjust sensitivity-decrease mV setting

    • C.pacingtreshold test-adjust output parameters

    • D.activate hysteresis


    15.What caused byis appropriate in this patient

    • A.adjust sensitivity-increase mV setting

    • B.adjust sensitivity-decrease mV setting

    • C.pacingtreshold test-adjust output parameters

    • D.activate hysteresis


    ad