Alternate Method to Certify Grid Plates and Line Scales using a CMM with Chromatic Confocal Sensor verses Traditional Methods. Hexagon Metrology By: JHorwell Sr. Metrologist July 2013. What may I Learn or Why Listen?.

loader
I am the owner, or an agent authorized to act on behalf of the owner, of the copyrighted work described.
capcha
Download Presentation

PowerPoint Slideshow about ' Hexagon Metrology' - huy


An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -
Presentation Transcript

Alternate Method to Certify Grid Plates and Line Scales using a CMM with Chromatic Confocal Sensor verses Traditional Methods

Hexagon Metrology

By: JHorwell Sr. MetrologistJuly 2013


What may i learn or why listen
What may I Learn or Why Listen? using a CMM with Chromatic Confocal Sensor verses Traditional Methods

  • Alternate method for certifying Line Scales

  • How to apply a white light confocal sensor for edge detection

  • Difference between displacement and intensity

  • Density of the data

  • Selection Criteria

2013 NCSLI International Workshop and Symposium


Agenda
AGENDA using a CMM with Chromatic Confocal Sensor verses Traditional Methods

01

02

03

04

Overview of Requirements

Brief History

Test Piece

Results

05

06

Error Budget

Lessons Learned

2013 NCSLI International Workshop and Symposium


Overview
Overview using a CMM with Chromatic Confocal Sensor verses Traditional Methods

  • Test the ability of a confocal chromatic sensor to find edges

  • Review existing Hexagon Hardware and firmware to accomplish this task

  • Compare distance measurements against certified artifact

  • Determine the feasibility to certify grid plates and glass line scales on a high accuracy CMM with confocal chromatic sensor

2013 NCSLI International Workshop and Symposium


History
History using a CMM with Chromatic Confocal Sensor verses Traditional Methods

  • Currently most line scales are measured on special purpose hardware consisting of an XY stage, Laser Interferometer feed back for position and an optical measuring device

  • Others use comparative techniques from a certified master scale to the scale under test

  • Some have combinations of manual and automatic systems

  • In the past some used high accuracy CMM with an image analysis system to certify grid plates

2013 NCSLI International Workshop and Symposium


200mm glass scale with chrome lines
200mm Glass Scale with Chrome Lines using a CMM with Chromatic Confocal Sensor verses Traditional Methods

  • Task is to be able to find the edges with a standard deviation less than 0.09um

  • Calculate the distances and compare these to a certified value.

2013 NCSLI International Workshop and Symposium


Equipment selection criteria
Equipment Selection Criteria using a CMM with Chromatic Confocal Sensor verses Traditional Methods

  • Small Measurement Uncertainty ≤ 0.3 + L/1000

  • Non Contact Sensor Availability

  • Stability over time

  • Hardware, Software and Firmware available and capable for finding edges

2013 NCSLI International Workshop and Symposium


Leitz pmm c 12107 infinity
Leitz PMM C 12107 Infinity using a CMM with Chromatic Confocal Sensor verses Traditional Methods

  • Measuring Error MPE in [µm] according to ISO 10360-2 (2010-6)

    • E0 ≤ 0.3 + L/1000

    • E150 ≤ 0.5 + L/1000

    • R0 ≤ 0.3

  • Scale Resolution

    • 4nm

  • LSP –S4 Probe head with integrated optional Precitec LR Sensor

  • 2013 NCSLI International Workshop and Symposium


    Precitec lr sensor fully integrated into leitz pmm c infinity
    Precitec LR Sensor Fully Integrated into Leitz PMM C Infinity

    • Numeric aperture 0.66

    • Measuring procedure chromatic confocal

    • Measurement angle relative to surface 90°±40°

    • Measuring distance 6.5 mm

    • Diameter of measuring spot 1.4 µm

    2013 NCSLI International Workshop and Symposium


    90 degree sensor
    90 degree Sensor Infinity

    • We are continually developing sensors and adaptors for our systems

    • The sensor to the right is similar to the sensor we used but can measure the line scales in the vertical direction verses horizontal

    • The sensor can be manually rotated and our controller can handle more than one sensor on a single system

    2013 NCSLI International Workshop and Symposium


    Overview glass line scales
    Overview Glass Line Scales Infinity

    • Find the edges of a Chrome Lines on Glass Line Scales using the Precitec LR Chromatic Confocal Sensor on the Leitz infinity 12107 in RI

      • Build a reference system on the Glass Scale using tactile probing

      • Build an alignment system using the Precitec LR Sensor

      • Measure each reference line via 3-pts and constructing an axis

      • Intersect each reference line with the alignment axis

      • Calculate the distance from each line to the zero line and compare with certificate of certification

      • Test 2-methods for finding the edge

      • Test alignment stability

      • Run repeat test to determine standard deviations

    2013 NCSLI International Workshop and Symposium


    Test part glass line scale
    Test Part – Glass Line Scale Infinity

    • Certificate of Calibration No 115-01538 by CIPM MRA

    • Uncertainty of Measurement: U = 0.10µm + L/2000

    • The distances between the graduations were measured with a photo mask measuring system consisting of a precision x-y bearing table, a two-axis differential plane mirror interferometer and a microscope. The table moves the graduations to the focus of microscope while the interferometer measures its position. The relative position of a graduation within the field of view the microscope is determined by digital image analysis.

    • The scale was cleaned, supported at the Bessel-points (distance 121mm) and measured with the graduation on top using episcopic white light illumination. For each graduation the position of the trailing edge was measured at the height of the collinear alignment marks

    2013 NCSLI International Workshop and Symposium


    What was missing from the hardware configuration
    What was missing from the hardware configuration Infinity

    • While we could find the edge of the lines using traditional analysis by automatically finding extreme points with a standard deviation between 112nm and 250nm using a 30 repeat test. We determined that these values were not as good as we believe our equipment was capable up.

    • Our Firmware Development Group made a flag within our firmware that allowed us to receive the intensity out verses displacement from the Precitec LR Sensor

    • This was the break thru that we needed to accomplish our task

    2013 NCSLI International Workshop and Symposium


    Alignment fixture procedure for glass scale
    Alignment & Fixture Procedure For Glass Scale Infinity

    • The Glass Scale was place on 3-spherical pins which were near the airy points with 2 at one level and a single pin at the other airy point.

    • 2-posts were used to align the scale along its axis and a single pin for the stop on the zero end plane of the Glass Scale

    • Tactile alignment

      • Using the LSP-S4 probe head and ruby sphere probe we measured a plane on the top and long side and a single point on the end face. This provided us with the rough alignment before we picked the Precitec LR Sensor from our Probe Change Rack

    • Optical Alignment

      • Scanned two lines on the top surface of the Glass Scale, constructed a plane and defined the Z-axis spatial vector and alignment from that.

      • We measured 3-points on the line between the 0 and 10mm lines and constructed axis

      • Measure same on the last line between the 190 and 200mm marks

      • Constructed an axis from these two points which define the rotational portion of our Coordinate system

      • Measured 3-points and constructed an axis on the back side of the zero line at ±0.5mm from the axis just created

      • Intersected this zero line axis with the line the defined the rotational portion of our Reference System and this became the X-zero of our Glass Scale

    2013 NCSLI International Workshop and Symposium


    Alignment of glass scale using precitec lr sensor

    Found 3-edge points at ±0.5mm from zero alignment axis on the 0-position vertical line and built an axis and the intersection of that axis and the zero alignment line created the XY origin

    0

    10

    190

    200

    Found 3-edge points at 0.5mm spacing on each of the 2-horizontal lines and constructed an axis between then which defined the axial alignment

    Alignment of Glass Scale using Precitec LR Sensor

    2013 NCSLI International Workshop and Symposium


    Measurement sequence
    Measurement Sequence the 0-position vertical line and built an axis and the intersection of that axis and the zero alignment line created the XY origin

    • We started our measuring sequence by measuring the Zero Line using 3-points; constructing an axis and intersecting the newly measured axis with the directional axis of our reference system and got a single point for each line measured.

    • We basically stepped and repeated this process until all lines were measured

    • We then calculated the distance in X-direction between each line measured and the Zero Line.

    • We repeated this process 10 times

    2013 NCSLI International Workshop and Symposium


    Finding the edge
    Finding the edge the 0-position vertical line and built an axis and the intersection of that axis and the zero alignment line created the XY origin

    • What are some of the issues with finding the real edge and then correlating this to traditional methods.

      • Edge is not perpendicular to the reference plane

      • Edge is not straight

      • Does back lighting verses top lighting give two different edges

      • Can we find the same edge as traditional optical systems and measuring microscopes

    2013 NCSLI International Workshop and Symposium


    Finding extreme points on the edge
    Finding extreme points on the edge the 0-position vertical line and built an axis and the intersection of that axis and the zero alignment line created the XY origin

    Edge Detail

    2013 NCSLI International Workshop and Symposium


    Summary of results
    Summary of Results the 0-position vertical line and built an axis and the intersection of that axis and the zero alignment line created the XY origin

    2013 NCSLI International Workshop and Symposium


    Current procedure
    Current procedure the 0-position vertical line and built an axis and the intersection of that axis and the zero alignment line created the XY origin

    • The 3 slides following this show the results from finding the edge of a single chrome line using ScanOnLine with the following point spacing:

      • 0.0001 = 10,000pts/mm

      • 0.00005 = 20,000pts/mm

      • 0.000025 = 40,000pts/mm

    • I am hoping to have even lower standard deviations that I originally obtain using the Precitec LR sensor on Infinity

    • Currently I am finding the edge by selecting points based on light intensity of 0.26 to 0.35

    • When I am on the chrome surface I get saturation which is near a value of 1

    2013 NCSLI International Workshop and Symposium


    Finding the edge alignment line
    Finding the Edge Alignment Line the 0-position vertical line and built an axis and the intersection of that axis and the zero alignment line created the XY origin

    • Top picture illustrates the edge as seen by the CMM system in red

    • The blue line is the change in intensity as the system crosses the edge

    • The small red points on top of the blue line are the points used to define the edge based on intensity value.

    • Lower picture shows the measured edge from the CMM in Red

    • The line in black shows the intensity curve from the Crocodile box. The change in intensity is from glass surface to the silver plated surface

    2013 NCSLI International Workshop and Symposium


    Density at 10 000pts mm
    Density at 10,000pts/mm the 0-position vertical line and built an axis and the intersection of that axis and the zero alignment line created the XY origin

    Chrome Surface

    Glass Surface

    Measured Curve

    Intensity Points Selected

    Scan Direction

    2013 NCSLI International Workshop and Symposium


    Results
    Results the 0-position vertical line and built an axis and the intersection of that axis and the zero alignment line created the XY origin

    • The red line represents average value from 10 repeats – the certified value. I did not apply temperature compensation at the moment but could and most likely will.

    • The Blue dots represent the results from the 25 runs

    2013 NCSLI International Workshop and Symposium


    Results numerical
    Results - numerical the 0-position vertical line and built an axis and the intersection of that axis and the zero alignment line created the XY origin

    2013 NCSLI International Workshop and Symposium


    Still not Satisfied! the 0-position vertical line and built an axis and the intersection of that axis and the zero alignment line created the XY origin

    Noise reduction algorithm affects the edge detection!

    2013 NCSLI International Workshop and Symposium


    Raw data from cmm with out noise reduction
    Raw Data From CMM with out Noise Reduction the 0-position vertical line and built an axis and the intersection of that axis and the zero alignment line created the XY origin

    2013 NCSLI International Workshop and Symposium


    Selection based on intensity data shown has no noise reduction

    1µm

    • The points inside this window represents original selection for intensity:

      • Green = 0.5 to 0.6 Intensity 0.26 to 0.35

    Selection based on Intensity Data Shown has NO Noise Reduction

    2013 NCSLI International Workshop and Symposium


    Selection based on intensity data shown has no noise reduction1
    Selection based on Intensity Data Shown has on intensity:NO Noise Reduction

    • The original testing presented in this paper used the 0.26 to 0.35 intensity selection criteria and noise reduced data from the CMM

    • We wanted to test raw data with no noise reduction and a steeper point on the intensity curve.

      • Selected 0.5 to 0.6 Intensity

      • Selected 0.5 to 0.71 Intensity

    • Results are taken weeks apart with completely new setups and the standard deviations for this newest procedure has a maximum σ of 0.55µm

    2013 NCSLI International Workshop and Symposium


    Intensity selection comparison
    Intensity Selection Comparison on intensity:

    2013 NCSLI International Workshop and Symposium


    Straightness plot of the scan yz plane
    Straightness Plot of the Scan (YZ plane) on intensity:

    2013 NCSLI International Workshop and Symposium


    Verification results are equal in the and scan direction
    Verification results are equal in the + and – Scan Direction

    2013 NCSLI International Workshop and Symposium


    Verification results are equal in the and scan direction1
    Verification results are equal in the + and – Scan Direction

    2013 NCSLI International Workshop and Symposium


    Optiv performance 2z443
    Optiv Direction Performance 2z443

    • Measuring Error MPE in [µm] according to ISO 10360-7

      • EX , EY ≤ 2.2 + L/125

    • Standard Scale Resolution

      • 0.5µm

  • Optional Scale Resolution

    • 0.1µm

  • 2013 NCSLI International Workshop and Symposium


    Optiv results
    OPTIV Results Direction

    [2.2 + L/125]µm

    2013 NCSLI International Workshop and Symposium


    Error budget
    Error Budget Direction

    • Environment 20°C±0.1°C 10%CTE of 8.3*10-6/°C = 0.016µm

    • Reproducibility – actual 25 repeats plus realignment = 0.083µm

    • CMM Accuracy Parallel to X extrapolated from physical measurements [0.2 + L/1500]µm

    • Other considerations

      • Quality of edge

      • Squareness of measured lines to center axis

    • UC = [0.3 + L/1000]µm with coverage factor of 2

    2013 NCSLI International Workshop and Symposium


    Summary
    Summary Direction

    • We wanted to show that one could calibrate a Glass Line Scale and or a grid plate via a non traditional method such as presented within this paper. We hope that the community will consider this alternate method and hope to engage in some round robin testing to verify that a high accuracy CMM with chromatic confocal sensor could be used and that is process could allow users to verify their working standards without the need for special purpose systems. It also adds validity for owning instruments of this caliber.

    • We are not claiming that this method is faster than a tradition non-contact CMM, but we do feel that we can reduce the measurement uncertainty for calibrating longer artifacts with a significant lower measurement uncertainty and without the need for a dedicated single purpose system.

    2013 NCSLI International Workshop and Symposium


    Lessons learned
    Lessons Learned Direction

    • It is possible to find an edge using a chromatic confocal sensor using only displacement

    • Light intensity changes are a better method to find the edges than using displacement values only

    • Quality of the edge has a significant impact to our standard deviation

    • Use of noise reduction algorithms has some impact on the standard deviation, but changing the selection criteria had very little impact

    • Correlation to alternate calibration methods are easy to reach with the Leitz Infinity with integrated Precitec LR chromatic confocal sensor

    2013 NCSLI International Workshop and Symposium


    The End Direction

    Thank you!