Pupil centre shift between bright light and dim light conditions

Pupil centre shift between bright light and dim light conditions. C. Winkler von Mohrenfels 1 , J Salgado 1, R. Khoramnia 1 , Christian Wüllner 2 , Peter Riedel 2 , C Donitzky 2 1Augenklinik, Klinikum rechts der Isar, Technische Universität München, Munich, Germany, Financial interest: none

hu-ball
Download Presentation

Pupil centre shift between bright light and dim light conditions

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Pupil centre shift between bright light and dim light conditions C. Winkler von Mohrenfels1, J Salgado1, R. Khoramnia1, Christian Wüllner2, Peter Riedel2, C Donitzky2 1Augenklinik, Klinikum rechts der Isar, Technische Universität München, Munich, Germany, Financial interest: none 2 WaveLight AG Erlangen, financial interest: employee

  2. Problem • different pupil size between dim and bright light condition • centration of laser ablation: pupil center, apex, line of sight • eye tracking: pupil, iris • pupil shift between the different light conditions could lead to miss alignment of the laser ablation. Method • 47 healthy subjects both eyes (12♂, 35 ♀), • mean age 36 years (range 23 to 60) • WaveLight Allegro Topolyzer: Pupil software (dim [20 lux] and bright light [455 lux] conditions)

  3. Pupil difference and age 5.0 4.5 4.0 3.5 3.0 Pupil devition between dim and britgh light conditions in [mm] 2.5 2.0 1.5 y = -0.0252x + 3.8457 R 2 = 0.2163 1.0 0.5 0.0 10 20 30 40 50 60 70 age in [years]

  4. Pupil center shift right eye x axis 0.5 0.4 0.3 0.2 0.1 temporal • y-axis nasal 0.0 -0.5 -0.4 -0.3 -0.2 -0.1 0.0 0.1 0.2 0.3 0.4 0.5 -0.1 -0.2 -0.3 -0.4 -0.5 MW 105µm (SD 65; min 18µm; max 330µm)

  5. Pupil center shift left eye x-axis 0.5 0.4 0.3 0.2 0.1 y- axis nasal temporal 0.0 -0.5 -0.4 -0.3 -0.2 -0.1 0.0 0.1 0.2 0.3 0.4 0.5 -0.1 -0.2 -0.3 -0.4 -0.5 MW 107µm (SD 64; min 11µm; max 348µm)

  6. Conclusions • the average pupil shift is 0.11mm • most pupils shift to temporal side • huge variation 0.01mm to 0.348mm • the larger pupils shifts could lead to a miss alignment between tracking and laser ablation. • → risk of decentration and poor visual outcome