An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. RAD 354 Chapt. 11 Control of Scatter • Break down into: Those that reduce pt. dose and those that are geometrical in nature and those NOT! • 3 factors affecting scatter (primary) • Increased kVp • Increased field size • Increased pt. thickness

  2. Spatial Resolution & Contrast Resolution • Spatial resolution may be thought of as GEPMETRIC in nature (F.S. size, emission spectrum, etc. dealing with geometric image formation • Contrast resolution – driven by scatter and other sources of “noise”

  3. Scatter • Increased field sizes = more scatter – collimation is the most readily available and easiest thing to lower the amount of scatter • Patient thickness also increases scatter- compression may be used to help avaid this (IVP’s and mammo are examples)

  4. Beam restricting devices limit the radiation to the patient • Aperaturediaphram (size and resultant filed size are a direct proportion – draw the damn picture and figure the problems) • Cones and cylinders – great for absorbing scatter, but are circular shaped = great for improving contrast and removing scatter BUT require MUCH MORE mAs as a result!

  5. Variable aperaturediaphram • Mandated in 1974 by the US Dept. of Food and Drug Admin. (mandate later removed) • Positive Beam Limitation Devices (PBL’s) • Automatically collimate to the size of the cassette/receptor in the bucky and CANNOT be a bigger size than the cassette/receptor

  6. Filtration/Grid • Filtration also will decrease the low energy rays and limit patient dose and some scatter • Grids remove “angled scatter” so only FORWARD scatter is of any benefit to the radiographic image – all other scatter degrades the image!

  7. Scatter = LOWER contrast • Using a grid (alternating strips of fine/thin leaded strips with alternating radiolucent interspace material) can effectively reduce the amount of ANGLED scatter from reaching the film/recepter

  8. Grid Terms • Grid Ratio = height of the lead lines divided by the interspace WIDTH • Grid frequency/lines per inch = the more lines per inch, the more clean up • Grid clean up = scatter w/o a grid vs scatter reaching film with a grid AKA “Contrast Improvement Factor” • Grid function = improve image contrast

  9. Bucky Factor • Refers to the AMOUNT of radiation to the patient with a grid vs W/O a grid • Higher the grid ratio, the higher the “bucky factor” • The higher the kVp, the higher the “bucky factor” Grid weight refers to how heavy it is – DUH! The more lead, the heavier it is!

  10. Grid Types • Parallel • Crossed (cross-hatch) • Focused • Focused - crossed

  11. Grid Problems • Grid cut-off = short SID’s result in the vertical, parallel strips absorbing the “diverging” beam at the outer margins of the grid/film/rescepter MOST pronounced at short SID’s • Most grid problems are “positioning related” • Uneven grid/off level grid • Off centered (lateral decentering) • Off focus grid • Upside down focused grid

  12. Focused Grid Misalignment • Off level = grid cutoff across image; underexposed image (light) • Off center = ditto • Off focus = CR centered to one side or the other of a focused grid • Upside down grid = severe grid cut-off )no density) at BOTH sides of the image

  13. Grid Ratio Selection • 8:1 grid is the most widely used • 5:1 grid most widely used PORTABLE grid • Grid ratio is kVp driven • Higher kVp’s warrant higher grid ratios • Higher grid ratios = higher patient dose (more radiation needed to produce an image) • As kVp increases past MAXIMUM OPTIMUM kVp, patient dose INCREASES

  14. mAs- Grid considerationsAs grid ratio increases, so must mAs! • 5:1 grid = 2 X mAs • 8:1 grid = 4 X mAs • 12:1 grid = 5 X mAs • 16:1 grid = 6 X mAs

  15. Air gap technique • By allowing the scatter radiation to “diffuse” in the atmosphere after the patient but BEFORE the film/receptor as the scatter diffuses and does NOT reach the film/receptor • C-spine lateral is a good example of this