1 / 22

ZOLL M-Series

Objectives. At the end of this module the participant can: Properly place of 3,5, or 12 lead ECG electrodes Print option for 3- or 5-lead Connect the ECG cable to the back of the machine . Monitoring the Patient . This module will address monitoring done with 3 lead, 5 lead and 12 lead electr

hayes
Download Presentation

ZOLL M-Series

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


    1. ZOLL M-Series/E-Series Biphasic: 3, 5, 12 Lead Monitoring Self-Paced Training Module #10

    2. Objectives At the end of this module the participant can: Properly place of 3,5, or 12 lead ECG electrodes Print option for 3- or 5-lead Connect the ECG cable to the back of the machine

    3. Monitoring the Patient This module will address monitoring done with 3 lead, 5 lead and 12 lead electrodes only. However, monitoring can be done with electrodes, Stat-Padz, or paddles. A review of patient preparation will be given for both 3, 5 and 12 lead monitoring, then the module will be separated in 2 sections: 3 and 5 lead monitoring 12 lead monitoring

    5. These illustrations demonstration ECG placement for the 3 and 5 lead cable and electrodes.

    6. Selector Switch Turn Selector Switch to Monitor (Gray) area Press the Lead button to scroll through the different leads available Press the Size button to increase the size of the ECG on the screen Press the Recorder button to start or stop printing and recording of the event on the screen

    7. The 3 or 5 lead cable will fit snugly into the back of the M-Series in the ECG port. The ECG cable should be clean, dry, and without any cracks.

    8. Electrode Placement 12-Lead Locate proper V1 site Find jugular notch and move finger down 1.5 inches to the Angle of Louis (where the manubrium joins the sternum) feels like a ridge. Locate fourth intercostal space on right side lateral below the Angle of Louis. This is the proper placement of the V1 lead.

    9. Electrode Placement 12-Lead After locating the V1 lead, the rest of the leads can be placed according to the following illustrations.

    10. Electrode Placement 12-Lead Locate Limb Lead Junction box in the middle of the ECG monitoring cable Remove protective cap from the connector labeled V1-V6 Place end of V lead cable into the connector To change from 12 lead to 4 lead Monitoring Disconnect V Lead cable from the V1-V6 connector in the middle of the ECG monitoring cable Place protective cap (attached to cable) back into connector

    11.

    12. Displaying the 12 Lead Menu Press the Lead button to display any lead selection except PADS Press the 12 lead softkey to access the menu

    13. The 12 Lead Menu Will Display at the Bottom of the Screen To acquire 12-lead waveforms, press Acquire softkey. ACQUIRING ECG message will appear. After 10 seconds, ACQUISITION COMPLETE message is displayed and 12-lead data automatically prints. Messages will alert practitioner if a problem such as Lead Off occurs Press HALT key (which becomes available) to abort the acquisition

    18. To enter patients age, press Age softkey Press Inc Age or Dec Age Press Enter softkey when selection is highlighted Default age is 41 To enter gender, press Gender softkey to toggle between male and female Press Enter once selection highlighted

    19. 12 lead

    21. If the M Series/E Series is configured to do so, the Transmission set-up menu automatically displays after 12 lead data has been acquired Faxing: Pre-Entered Phone # Press Prev Phone # or Next Phone # to scroll through options Press Dial Phone # M-series displays messages on the status of the fax

    22. Faxing: Dialing Manually Press Manual Dial softkey Press Inc digit or Dec digit to change the digit Press next digit to select next digit Repeat previous steps until phone # is entered. Press Dial

    23. This concludes Module #10 3, 5, and 12 Lead Monitoring Please continue on with other modules which may be required at your facility