Development of a Theoretical Model for Membrane Hydrophone Transfer Characteristics

Development of a Theoretical Model for Membrane Hydrophone Transfer Characteristics. Pierre G é lat. 3 April 2003. Background. Bilaminar hydrophones are extensively used to measure and characterise medical ultrasound fields

gloria
Download Presentation

Development of a Theoretical Model for Membrane Hydrophone Transfer Characteristics

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Development of a Theoretical Model for Membrane Hydrophone Transfer Characteristics Pierre Gélat 3 April 2003

  2. Background • Bilaminar hydrophones are extensively used to measure and characterise medical ultrasound fields • Developing a hydrophone model would enable users to improve the accuracy of their measurements in the field of medical ultrasound (phase response, deconvolution) • Type of hydrophone initially modelled: 25 m film thickness 0.5 mm diameter bilaminar membrane hydrophones • Need to generalise model to simulate the response of the new generation of Precision Acoustics bilaminar membrane hydrophones (15 m film thickness, 0.4 mm diameter)

  3. A bilaminar membrane hydrophone

  4. Diagram of a bilaminar membrane hydrophone

  5. Hydrophone model assumptions • 25 m pvdf film thickness, 0.5 mm diameter bilaminar • Assume plane wave incidence on hydrophone • Use transfer matrix formulation to evaluate pressure and particle velocities at extremities of pvdf layers • Account for secondary piezoelectric effects • Model pvdf layers as equivalent voltage sources in series with capacitance • Model leg as distributed series resistance shunt capacitance network • Model coaxial cable as transmission line • Account for electrical loading on cable (e.g. oscilloscope, amplifier, arbitrary load)

  6. Bilaminar Hydrophone Equivalent Electrical Circuit

  7. Model input parameters

  8. Open-circuit sensitivity

  9. Predicted phase response

  10. Prediction of pressure waveform

  11. Conclusions • Validation of a model for predicting the transfer characteristics of membrane hydrophones • Can be used to estimate true pressure waveform