An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Reference Membrane Location

  2. Reference Membrane

  3. Span Check • Perform 8 minute count • Compare mass value with ABS value • ABS must be within 5% or BAM will log CAL error (D error code)

  4. CALIBRATION MODE REF MBRN: < COUNT (Io): 634000 COUNT (I): 556234 CAL MASS M: 0.801 mg/cm2 START STOP EXIT Calibration Test Screen

  5. Hands On Workshop – Reference Membrane • Remove Reference Membrane • Inspect Reference Membrane • Reinstall Reference Membrane • Inspect mechanical movement of Reference Membrane • Perform Reference Membrane Count/Mass Test in TEST/CALIBRATE Screen • Compare Mass value with ABS value