1 / 11

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY

IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology. RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. Part 19.05: Optimization of protection in Mammography Practical exercise. Overview.

Download Presentation

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION INDIAGNOSTIC ANDINTERVENTIONAL RADIOLOGY Part 19.05: Optimization of protection in Mammography Practical exercise

  2. Overview • To be able to apply quality control protocol to mammography equipment • To measure the Half Value Layer • Interpretation of results 19.05 : Optimization of protection in Mammography

  3. IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology Part 19.05: Optimization of protection in Mammography Beam quality (HVL)

  4. Half value layer (HVL) • The Half Value Layer (HVL) can be assessed by adding thin aluminium (Al) filters to the X-ray beam and measuring the attenuation • Position the detector on top of the breast table • Place the compression device halfway between focal spot and detector 19.05 : Optimization of protection in Mammography

  5. Half value layer (HVL) • Select 28 kV and an mAs to produce at least 10 mGy and make an exposure • Position the aluminium filters on top of the compression paddle and assure that they intercept the entire radiation field. • Use the same mAs setting and make an additional exposure after adding each filter 19.05 : Optimization of protection in Mammography

  6. HVL Measurement Geometry X-Ray Tube Aluminium filter Diaphragm Compression paddle Detector Lead Breast support ~ 300 mm ~ 300 mm 19.05 : Optimization of protection in Mammography

  7. Half value layer (HVL) • For higher accuracy (about 2%) a diaphragm, positioned on the compression paddle, limiting the exposure to the area of the detector may be used • The HVL is calculated by applying the formula: 19.05 : Optimization of protection in Mammography

  8. Half value layer (HVL) Y0 :the direct exposure reading (mGy) Y1 and Y2 : the exposure with added aluminium thickness of X1 and X2 respectively Note 1:The purity of the aluminium must be 99.0% or greater. The thickness of the aluminium sheets should be measured to an accuracy of 1% 19.05 : Optimization of protection in Mammography

  9. Half value layer (HVL) Note 2:For this measurement the output of the X-ray machine must be stable Note 3:The HVL for other (clinical) energies, and other target materials and filters should also be measured for assessment of the average glandular dose 19.05 : Optimization of protection in Mammography

  10. Half value layer (HVL) Limiting value :For 28 kV Mo-Mo the HVL must be over 0.30 mm Al equivalent Frequency :Annually Equipment :Dosimeter, 99.0% aluminium sheets 0.20 and 0.40 mm 19.05 : Optimization of protection in Mammography

  11. Where to Get More Information European protocol for the quality control of the physical and technical aspects of mammography screening. http://euref.org/index.php?option=com_phocadownload&view=category&id=1&Itemid=8 American College of Radiology Mammography Quality Control Manual, Reston VA, 1999. 15.3: Optimization of protection in radiography

More Related