Advertisement
1 / 88

Renatron® II Service Maintenance Seminar PowerPoint PPT Presentation


  • 524 Views
  • Uploaded on 29-04-2012
  • Presentation posted in: General

Renatron® II Service Maintenance Seminar. Class Schedule. DAY ONE 8:30 AM- Seminar start: Introduction 10:00 AM- Break 12:00 PM- Lunch 2:30 PM - Break 5:00 PM- End of first day. Class Schedule. DAY TWO 8:30 AM- Seminar start 10:00 AM- Break 12:00 PM- Lunch 2:30 PM - Break - PowerPoint PPT Presentation

Download Presentation

Renatron® II Service Maintenance Seminar

An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -

Presentation Transcript


Renatron ii service maintenance seminar l.jpg

Renatron® II Service Maintenance Seminar


Class schedule l.jpg

Class Schedule

DAY ONE

8:30 AM- Seminar start: Introduction

10:00 AM- Break

12:00 PM- Lunch

2:30 PM - Break

5:00 PM- End of first day


Class schedule3 l.jpg

Class Schedule

DAY TWO

8:30 AM- Seminar start

10:00 AM- Break

12:00 PM- Lunch

2:30 PM - Break

3:30 PM- End of Seminar


Cautions l.jpg

Cautions

The Renatron® 100 is intended only for use with Renalin® 100 Cold Sterilant

Used only for Hemodialyzers that have been labeled multiple use.

Never run the machine without the uptake tube assembly

All hoses and lines for the Renatron® are cut to length and should not be altered.


Indications l.jpg

Indications

The Renatron® Dialyzer Reprocessing System is intended for in vitro, rinsing, cleaning testing, and delivery of proportioned Renalin® solution to hollow fiber dialyzers that have been validated to be reprocessed with the Renatron® II System and Renalin® sterilant. It is also indicated for record keeping of these functions.


Specifications l.jpg

Specifications


Electrical leakage current l.jpg

Electrical Leakage Current

100 micro amps

Selectable voltage 100, 120, 230, 240 VAC

The supply line voltage must not fluctuate more than 10% of the nominal voltage.

Fuses 100-120vac, one 4A

Fuses 220-240vac, two T2A


Environmental limits l.jpg

Environmental Limits

Operation Ambient Temperature Range

41-104° F (5-40° C)


Water requirements l.jpg

Water Requirements

Water must be filtered to particle size no larger than 1 micron.

Inlet Water Must Meet or Exceed AAMI Guidelines


Water temperature l.jpg

Water Temperature

Must be 59 -75°F (15-24° C)

A maximum temperature of 86° F is allowable if the dialyzer is stored immediately after reprocessing in an area where the temperature is between 59-75° F. (Must have letter)


Feed line flow l.jpg

Feed Line Flow

1.725 liters (.46 gallons) per minute average

Maximum water pressure is 55 PSI static

Minimum water pressure is 20 PSI dynamic read in step 4.

Optimum pressure of 30-35 PSI in Step 4.


Peak flow rate l.jpg

Peak Flow Rate

6 liters (1.6 gallons) per minute

15-35 seconds in duration

Occurs at approximately 3 minute intervals

3 times per cycle


Drain requirements l.jpg

Drain Requirements

Must be able to accept 6 liters (1.6 Gallons) per minute per Renatron®.


Materials of construction l.jpg

Materials of Construction

The drain should be constructed of PVC/CPVC or other Non-corrosiveMaterial


Renatron ii station components l.jpg

Renatron II Station Components


Renatron ii station components16 l.jpg

Renatron® II Station Components

Renatron® II Dialyzer Reprocessing Station

Has a hydraulic side and electronic side.

1 - 6 ft water supply hose with back flow preventor.

1 - 6 ft drain hose.

1 - 6 ft Renalin® hose.

1 - 6 ft RS-232 communication cable


Renatron ii station components17 l.jpg

Renatron® II Station Components

  • Pressure gauge block/regulator

  • 70 cc cal cell

  • Drip tray-single or dual

  • Renalin®100 Uptake tube assembly


Renatron ii station components18 l.jpg

Renatron® II Station Components

  • Computer

  • Top of the line computer system

  • Windows XP

    Monitor

  • VGA Monitor- CRT or Flat Screen


Renatron ii station components19 l.jpg

Renatron® II Station Components

  • Renalog RM Data Management System

  • Dedicated label printer

  • Dedicated Report Printer

  • Interface Module

  • Multiplexing module that can connect up to 6 Renatrons®.

  • Renalog RM – 12 Renatrons per computer


Overview of dialyzer reprocessing l.jpg

Overview of Dialyzer Reprocessing


End of treatment l.jpg

End of Treatment

After treatment is completed, the dialyzer is brought to the reuse room.

The time interval between the end of the treatment and reprocessing the dialyzer should be kept within the time limits prescribed by the physician

The suggested maximum time interval is two hours


Dialyser cleaning l.jpg

Dialyser Cleaning

Many clinics will clean or rinse a dialyzer before reprocessing.

Blood side flush

RUF

Renaclear®


Reprocessing l.jpg

Reprocessing

  • Attaching a dialyzer

    Connect:

    • Venous blood side

    • Venous dialysate side

    • Arterial dialysate side

    • Arterial blood side


End of reprocessing cycle l.jpg

End of Reprocessing Cycle

Make sure reset is pressed before the dialyzer is disconnected.

Disconnect the dialyzer in the reverse order that it was connected. Cap each port as soon as it is disconnected from the Renatron®.

Aseptic technique must be followed while removing the reprocessed dialyzer from the Renatron® station to prevent contamination of the dialyzer blood pathway


Dialyzer storage l.jpg

Dialyzer Storage

Storage Area

Well ventilated

Temp. range 59-75°F

Out of direct Sun light

Dialyzer Storage

Minimum storage time is 11 hours

Maximum storage time ?

Temperature differences may cause pressure buildup in dialyzer


Dialyzer testing before treatment l.jpg

Dialyzer Testing before Treatment

The dialyzer must be tested for Renalin® 100 content after proper storage time- Perrassay Test Strip

The dialyzer is rinsed on the dialysis machine and a Residual Test Strip is used to verify that the Renalin® 100 content has been rinsed to acceptable levels

Recirculate until treatment is started to prevent rebound


Renatron ii system operation l.jpg

RENATRON® II SYSTEMOPERATION


Power up sequence l.jpg

Power Up Sequence

Power on sequence

Printers

Monitor

Computer

Interface module

Renatron® II

All at once if using the power strip


Renalog rm start up l.jpg

Renalog® RM Start Up

  • Log in

  • User ID

  • Password

  • Mode

  • Change Time- Activation key needed

  • Start Up Screens

  • Renalin Verifications

  • Service Rinse

  • Sanitize Renatrons

  • Calibration

  • Water Verifications- 20 psi to 55 psi

  • Water Cultures

  • Dialyzer Verifications- Precleaned, Expired, and not Processed


Renatron ii 100 startup l.jpg

Renatron® II 100 Startup

Renalog® RM Startup

Renalin® 100 is connected

The Renatron® II 100 is sanitized

The Renatron® II 100 is calibrated

Static and dynamic water pressure readings


Renatron ii 100 startup31 l.jpg

Renatron® II 100 Startup

Calibration verification

Scan barcodes- Authentication label, Renatron, and Dialyzer

The Renatron® II 100 runs to step 57

Calibration information is stored in the computer and a calibration report can be generated.


Modes of operation l.jpg

Modes of Operation


Program mode l.jpg

Program Mode

Low flux dialyzers (00) kuf <8. Program time = 8 minutes.

High efficiency dialyzers (CH) kuf 8-15. Program time= 10 minutes.

High flux dialyzers (HF) kuf 15+. Program time = 10 minutes.


Automatic reprocessing l.jpg

Automatic Reprocessing

Scan Authentication label, dialyzer barcode label, and Renatron® barcode.

Minimum Volume

Program Mode

When reset is pressed in step 57, labels are printed and the database is updated.


Semi automatic l.jpg

Semi-Automatic

Click on F6 to display Patient list

Select patient’s dialyzer

Select type of reprocessing

Preclean

Reprocess without a use

Reprocess after use

Sets the Renatron® II 100 to the proper settings for that dialyzer


Manual reprocessing with computer l.jpg

Manual ReprocessingWith Computer

Process dialyzer on Renatron® II 100

Set Hold to Set volume to dialyzer minimum volume setting

Press mute and reset button to select program setting

At end of cycle, enter information into computer

Preprocess

Reprocess without use

Reprocess after use

Renatron number

Container number

Current Volume

Did the dialyzer fail?


Manual reprocessing with renatron ii 100 only l.jpg

Manual Reprocessingwith Renatron® II 100 only

Process dialyzer on Renatron® II 100

Set Hold to Set volume to dialyzer minimum volume

Press mute and reset button to select program setting

At end of cycle, hand write label


Preclean mode l.jpg

Preclean Mode


Pre clean mode l.jpg

Pre-clean Mode

  • Pre-cleaning a dialyzer will help breakdown residual blood in the dialyzer.

  • Pre-clean program works for all types of dialyzers. Program time  4 minutes.

  • Recommended contact time for pre-cleaned dialyzers is 10 minutes to two hours.


Caution l.jpg

Caution

Pressure buildup may occur in dialyzers left on the Renatron®

Take care when removing the dialyzer

Pressure buildup may cause fluid to spray


Barcode labels l.jpg

User Authentication Label

Information Label

Barcode labels

Dialyzer Barcode Label

Pre-Cleaned Label

Calibration Cell Label

Failed Label


Sanitize program l.jpg

Sanitize Program


Daily sanitize l.jpg

Daily Sanitize

  • The Renatron needs to be sanitized once a day.

  • The machine must be connected in sanitize configuration. “SEL” must be in the top display and any of the programs “00”, “CH” or “HF” in the bottom display before “Start Sanitize” is pressed

  • The machine will display “sanitize complete” and a 3 second tone will sound at the end of step 83.

  • Turn off the Renatron and the water supply to the Renatron.


Minimum sanitize time l.jpg

Minimum Sanitize Time

  • The Renatron must sit for a minimum of 6 hours before it can be used.


Maximum sanitize time l.jpg

Maximum Sanitize Time

The Renatron® can sit for a maximum of seven days before it needs to be resanitized.


10 minute sanitize cycle l.jpg

10 Minute Sanitize Cycle

  • Press start process at the end of step 83, after a 10 minute hold in step 84, the machine will run to step 97, and signal sanitize complete. All fluid will be removed from the machine.

  • This cycle is to be used after a 409® procedure, when maintenance is performed on the machine, or when the machine is being returned for maintenance.


External disinfection daily l.jpg

External Disinfection(Daily)

1 % Renalin® solution is used to wipe down external surfaces of the Renatron® station

Actril® undiluted is also acceptable


Formula 409 cleaning l.jpg

Formula 409 Cleaning


409 cleaning l.jpg

409® Cleaning

Institutional Formula 409®

Institutional Formula 409® Cleaner Degreaser only

Must say NQF

Do not use any other solutions

Internal fluid pathways must be cleaned periodically ( typically once every two weeks )

This should be performed on Saturday a long dwell time is preferred


409 cleaning50 l.jpg

409® Cleaning

Institutional Formula 409® is free when Renalin® 100 is ordered in proper quantities


Dialyzer reprocessing cycle l.jpg

Dialyzer Reprocessing Cycle


Cleaning cycle step 8 to 24 l.jpg

Cleaning CycleStep 8 to 24

Chemical Draw- Step 9

Flush Blood Circuit Arterial to Venous-Step 14

Reverse Ultrafiltration- Steps 16 and 18

Flush Dialysate Compartment- Step 20

Flush Blood Circuit (Arterial to Venous)-Step 21

Flush Blood Circuit (Venous to Arterial)-Step 25


Test step 25 to 40 blood compartment volume measurement step 25 to 35 l.jpg

Test- Step 25 to 40Blood Compartment Volume MeasurementStep 25 to 35

Fill Blood Compartment from venous to arterial- Step 25

Sweep Blood Compartment- Steps 29, 31 and 33

Measure Mass- Step 35

Repeatability ± 5%


Test step 25 to 40 pressure leak test steps 36 to 39 l.jpg

Test- Step 25 to 40Pressure Leak Test- Steps 36 to 39

Draw vacuum- Steps 36 and 38

Measure bleed off rate- Step 39

Dialysate compartment is at -250 mmHg or less

Blood compartment atmospheric pressure

Leakage rate of .83mmHg per second fails ± 10 % for 00, CH mode

Leakage rate of 1.25mmHg per second fails ± 10% for HF program


Disinfect steps 41 to 57 l.jpg

Disinfect- Steps 41 to 57

Draw Renalin® - Step 45

Verify Renalin® Draw- Steps 46 and 47

Proportion Renalin® solution- Step 48

Flush Blood Compartment- Step 50

Flush Dialysate Compartment- Step 51

Draw Renalin® Solution to Dialysate Compartment- Step 53

Draw Renalin®Solution to Blood Compartment- Step 54


Interface module scanner operation l.jpg

Interface Module/Scanner Operation


Interface module l.jpg

Interface Module

  • Yellow power indicator light.

  • Green On Line light- Indicates that the computer is in communication mode. This light will only be on in the Automatic Reprocessing, Automatic Calibrations, or Renalin Connection screens.

  • On/Off switch and fuse at power entry module.

  • Older version of scanner will beep twice when scanning.

  • The newer version of the scanner will beep once when scanning.


Scanner l.jpg

Scanner

  • For best operation of Scan Team 3000 and 3400 models, hold the scanner so it is touching the barcode and press the trigger. Continue to press the trigger while moving the scanner away form the barcode until the scanner beeps.

  • Scan Team 3800 models, hold the scanner 3-9 inches from the barcode.

  • Make sure scanner lens is clean.

  • Clean the scanner with a nonabrasive cloth. Do not use a cleaner that contains alcohol.


Renalog rm label printer l.jpg

Renalog® RM Label Printer


Label printer l.jpg

Label Printer

  • Should be connected to the computer with a Parallel Printer Cable.

  • If printing is not aligned on label, perform the Calibration Procedure.

    • Turn power switch to Off

    • Hold Green Button down

    • Turn Power switch to On

    • Let go of Green Button when the red light flashes

    • Press Green Button to get “out of dump” label

  • Red light means that:

    • Cover is not completely closed

    • Ribbon is not advancing

    • Labels are not advancing


  • Renatron ii 100 controls and indicators l.jpg

    Renatron® II 100 Controls and Indicators


    Description of controls l.jpg

    Description of Controls

    • On

    • Off

    • Hold to Set

      • Set Knob

    • Start Process

    • Start Sanitize

    • Mute

    • Reset

    • Display


    Indicators l.jpg

    Indicators

    • Volume Display

    • Program Step

    • Clean

    • Test

    • Disinfect

    • Process Complete

    • Sanitize Complete


    Error messages l.jpg

    Error Messages

    • Volume Fail- Step 35

    • Pressure Fail- Steps 36, 38, or 39

    • Add Chemical- Steps 9, 45, 67, or 77

    • Tank Volume- Many steps throughout the program

    • Alarm

    • Interlock


    Safety features l.jpg

    Safety Features

    Tracks Preclean Dialyzers

    Tracks Maximum storage time

    Tracks Dialyzers that have not been processed- Preprocessed and Dry Packs

    Information stored on hard drive

    Precautions for similar name patients

    Patient data and program settings are password protected


    Renatron storage l.jpg

    Renatron® Storage

    Long term storage

    run machine through step 97 to ensure fluid pathways are drained

    Box in Renatron® box

    Keep in low traffic area


    Technical service l.jpg

    Technical Service

    Telephone Support

    24 hours a day 7 days a week

    1-800-328-3324

    Press “2” for “Medical Equipment Emergency” to get a live person

    Service

    Loaners available upon request

    Board exchanges for $200

    RMA’s must accompany all returns


    Warranty promotions l.jpg

    Warranty / Promotions

    Warranty

    1 year warranty from date of install

    90 days for service warranty

    90 days for parts orders

    Promotions

    PM contracts

    Service contracts


    Renalin 100 l.jpg

    Renalin® 100


    Warnings l.jpg

    Warnings

    • Corrosive.

    • Use universal precautions.

    • Can cause eye damage and skin irritation.

    • When handling Renalin®100, wear eye protection, rubber gloves and protective clothing.

    • Do not mix Renalin®100 with alkaline substances such as bleach.


    Indications71 l.jpg

    Indications

    Renalin® 100 Cold Sterilant is indicated for the in vitro cleaning and sterilizing of hollow fiber dialyzers with the RS8300 and RS8330 models of the Renatron® Dialyzer Reprocessing System labeled for use with Renalin®100. Renalin®100 Cold Sterilant also may be used for disinfecting dialysis equipment (e.g. kidney machines) and supplies (e.g. port caps).


    Contraindications l.jpg

    Contraindications

    Renalin® 100 Cold Sterilant is not designed, sold, or intended for use except as indicated.

    Patients with known hypersensitivity to hydrogen peroxide and/or peroxyacetic acid should not be treated using dialyzers reprocessed with Renalin® 100 Cold Sterilant.


    Water requirements73 l.jpg

    Water Requirements

    • AAMI quality water filtered to particle size no larger than 1 micron

    • Water temperature should be between

      59-75° F

    • Water Temperature may be up to 86° F if the dialyzers are stored at the proper temperature


    Renalin 100 storage l.jpg

    Renalin®100 Storage

    • Renalin®100 Storage

      • Upright

      • Never in direct sunlight

      • At temperatures between 32-75° F

      • Do not freeze

    • Never tamper with the cap

    • Never replace the vented cap with a nonvented cap

    • Renalin® can be removed from the shipping carton if it is not stored in direct sunlight


    Renalin 10075 l.jpg

    Renalin® 100

    • Renalin® 100 replaces

      • Formaldehyde

      • Bleach and/or hydrogen peroxide

    • Active ingredients

      • Hydrogen peroxide

      • Peroxyacetic acid

      • Acetic acid

    • Renalin® 100 degrades quickly and will not harm the environment. It breaks down into ; oxygen, water, and acetic acid.


    Renalin 10076 l.jpg

    Renalin 100®

    • Renalin® 100 is supplied in three different volumes:

      • Quart bottles

      • 2.5 gallon bottles that contain 2 liters of Renalin® concentrate

      • 1 gallon bottles containing 3 liters of Renalin® 100


    Renalin nomenclature l.jpg

    Renalin® Nomenclature

    • Renalin® 100 Cold Sterilant is a 100% Renalin® Solution.

    • Proportioned Renalin® - A 3.25% or 3.5% Renalin® solution that is used to sterilize the reprocessed dialyzer.

    • 1% Renalin® solution- A solution made from 1 part Renalin® concentrate and 99 parts AAMI quality water.


    Renalin nomenclature78 l.jpg

    Renalin® Nomenclature

    • 4% Renalin® solution is used to sanitize the Renatron®

    • 2% Renalin® solution is used during the preclean program


    Auto dilutor l.jpg

    Auto Dilutor

    The On Demand™ Auto Dilution System proportions Renalin® 100 Cold Sterilant and dispenses a continuous supply of 1% Renalin solution.


    Shelf life l.jpg

    Shelf Life

    • Renalin® 100 has a shelf life of 1 year from date of manufacture when properly stored.

    • Following dilution, Renalin® 100 exhibits a gradual loss of potency at a rate such that 50% of the active ingredients remain after a seven day period


    Shelf life81 l.jpg

    Shelf Life

    When diluting Renalin® Cold Sterilant to a 1% solution (e.g. port cap disinfection) fresh Renalin® Cold Sterilant should be used each time. Do not allow 1% Renalin® solution to sit for longer than 24 hours.


    Test strips l.jpg

    Test Strips

    • Perassay Test Strip

      • Simple go/no go test

      • Indicates a 1% Renalin® solution

      • Green stick for identification

      • Do not need to be refrigerated


    Test strips83 l.jpg

    Test Strips

    • Residual Test Strip

      • Color chart for interpreting test results

      • Do not need to be refrigerated

      • Verifies the proportioned Renalin® solution has been properly rinsed from the dialyzer prior to clinical use


    Renalin 100 presence test perassay l.jpg

    Renalin® 100 Presence Test(Perassay)

    • Remove 1 ml of solution from the dialysate port of the dialyzer

    • Immerse the entire reaction pad of the Perassay strip in the solution

    • Shake off excess solution

    • Read the reaction pad after 2-5 seconds. The test strip should turn a blue/gray or blue/black color. This will indicate 500 PPM of peracetic acid 1% Renalin® solution.


    Renalin 100 residual test strips l.jpg

    Renalin® 100 Residual Test Strips

    • Turn off the blood pump and clamp the arterial and venous blood lines adjacent to the access ends. Separate the blood lines where the access ends are joined and unclamp the venous blood line.

    • Allow a few drops of saline to drip from the venous line onto the residual test strip.

    • After 5 seconds, shake the excess liquid from the reaction zone and compare with the color scale on the container.

    • Per AAMI standards, a reading of less than 3 PPM is acceptable


    Handling and storage of reprocessed dialyzers l.jpg

    Handling and Storage of Reprocessed Dialyzers

    • If a dialyzer is foaming after being reprocessed it should be reprocessed again without a use

    • Store a reprocessed dialyzer in a cool, well ventilated area that is shielded from any direct sunlight. Recommended temperature range is 59-75° F

    • A reprocessed dialyzer should not be used until a minimum of 11 hours has elapsed

    • After storage and before rinsing, both headers of the dialyzers must be at least 2/3 full


    Renalin spill l.jpg

    Renalin® Spill

    • Flush spill with large quantities of water and also consult the handout “What to do if Renalin® spills”

    • After hours, contact Chemtrec for emergency spill information at

      800-424-9300 (202-483-7616 in the Washington D.C. area)


    Questions l.jpg

    Questions