Real-Time MRI – Outline

Real-Time MRI – Outline. Technical Considerations - Data Acquisition - Image Reconstruction Preliminary Applications - Joint Movements, Swallowing, Speaking - Heart Function A First Extension - Cardiovascular Blood Flow 3D MRI Movies ?!

enrico
Download Presentation

Real-Time MRI – Outline

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Real-Time MRI – Outline Technical Considerations - Data Acquisition - Image Reconstruction Preliminary Applications - Joint Movements, Swallowing, Speaking - Heart Function A First Extension - Cardiovascular Blood Flow 3D MRI Movies ?! - An (Almost) Unbelievable Story … Biomed NMR JF11/59

  2. Real-Time MRI – Robust, Fast, and Continuous Echo-Planar and Spiral Imaging - Limited Spatial Resolution (single-shot version) - Not Continuous (unless gradient-echo version with low flip angle) - T2* and Off-Resonance Problems Fast Spin-Echo (RARE) MRI - High SAR (high field), Not Continuous STEAM MRI - Low SNR, Not Continuous FLASH MRI - Very Short TR/TE (high bandwidth), Low SAR, Continuous - Fast? Undersampling! Biomed NMR JF11/52

  3. Real-Time MRI – Radial vs Cartesian Encoding Cartesian Simple Image Reconstruction - Inverse Fourier Transformation Radial Robust Against Motion - No Phase Encoding - Central Sampling of k Space Equivalence of Individual Spokes - Tolerance Against Undersampling - Minor Loss of Spatial Resolution Biomed NMR JF11/42

  4. Parallel MRI as Regularized Nonlinear Inverse Problem M Uecker et al, MagnReson Med 60:674-682, 2008;MagnReson Med 63:1456-1462, 2010 Biomed NMR JF11/65

  5. Real-Time MRI – Outline Technical Considerations - Data Acquisition - Image Reconstruction Preliminary Applications - Joint Movements, Swallowing, Speaking - Heart Function A First Extension - CardiovascularBlood Flow 3D MRI Movies ?! - An (Almost) Unbelievable Story … Biomed NMR JF11/59

  6. Real-Time MRI of the Knee Reconstructions Using Nonlinear Inversion Lifting of the Lower Leg in Prone Position Central Section Lateral Section Refocused Radial FLASH, TR/TE = 4.4/2.3 ms, 25o , 0.75 × 0.75 × 5 mm3 75 Spokes, 256 × 256 Matrix, Acquisition Time 333 ms, 3 Frames per Second Biomed NMR JF11/51

  7. Real-Time MRI of the Temporomandibular Joint Reconstructions Using Nonlinear Inversion Voluntary Opening and Closing of the Mouth Refocused Radial FLASH, TR/TE = 4.3/2.2 ms, 20o , 0.75 × 0.75 × 5 mm3 77 Spokes, 256 × 256 Matrix, Acquisition Time 333 ms, 3 Frames per Second Collaboration: N GersdorffBiomed NMR JF11/44

  8. Real-Time MRI of Normal Swallowing Reconstructions Using Nonlinear Inversion Voluntary Swallowing of 5 ml Pineapple Juice RF-Spoiled Radial FLASH, TR/TE = 2.17/1.44 ms, 5o, 1.5 × 1.5 × 10 mm3, FOV 192 × 192 mm2 19 Spokes, 128 × 128 Matrix, Acquisition Time 41.2 ms, 24 Frames per Second S Zhang et al, 2011 Biomed NMR JF11/46

  9. Real-Time MRI of Speaking – Sentences Reconstructions Using Nonlinear Inversion “You are watching a presentation about MRI in real time …” RF-Spoiled Radial FLASH, TR/TE = 2.22/1.44 ms, 5o, 1.5 × 1.5 × 10 mm3, FOV 192 × 192 mm2 15 Spokes, 128 × 128 Matrix, Acquisition Time 33 ms, 30 Frames per Second A Niebergall et al, 2011 Biomed NMR JF11/45

  10. Real-Time MRI of the Heart – Short-Axis View Reconstructions Using Nonlinear Inversion 15 Spokes 11 Spokes 30 ms 22 ms RF-Spoiled Radial FLASH, TR/TE = 2.0/1.3 ms, 8o, 2 × 2 × 8 mm3, FOV 256 × 256 mm2 15 / 11 Spokes, 128 × 128 Matrix, Acquisition Time 30 / 22 ms, 33 / 45 Frames per Second M Uecker et al, NMR Biomed 23:986-994, 2010 Biomed NMR JF11/53

  11. Real-Time MRI of the Heart – Short-Axis View Reconstructions Using Nonlinear Inversion 9 Spokes 7 Spokes 18 ms 14 ms RF-Spoiled Radial FLASH, TR/TE = 2.0/1.3 ms, 8o, 2 × 2 × 8 mm3, FOV 256 × 256 mm2 9 / 7 Spokes, 128 × 128 Matrix, Acquisition Time 30 / 22 ms, 55 / 71 Frames per Second M Uecker et al, NMR Biomed 23:986-994, 2010 Biomed NMR JF11/54

  12. Real-Time MRI of the Heart – 100 Frames per Second Reconstructions Using Nonlinear Inversion 5 Spokes 5 Spokes Short-Axis View Three-Chamber View RF-Spoiled Radial FLASH, TR/TE = 2.0/1.3 ms, 8o, 2 × 2 × 8 mm3, FOV 256 × 256 mm2 5 Spokes, 128 × 128 Matrix, Acquisition Time 10 ms, 100 Frames per Second M Uecker et al, NMR Biomed 23:986-994, 2010 Biomed NMR JF11/55

  13. Real-Time MRI of the Heart – 1.5 mm Resolution Reconstructions Using Nonlinear Inversion 15 Spokes, 34 ms 15 Spokes, 34 ms Short-Axis View Three-Chamber View RF-Spoiled Radial FLASH, TR/TE = 2.28/1.46 ms, 8o, 1.5 × 1.5 × 6 mm3, FOV 256 × 256 mm2 15 Spokes, 176 × 176 Matrix, Acquisition Time 34 ms, 30 Frames per Second M Uecker et al, NMR Biomed 23:986-994, 2010 Biomed NMR JF11/56

  14. Real-Time MRI – Outline Technical Considerations - Data Acquisition - Image Reconstruction Preliminary Applications - Joint Movements, Swallowing, Speaking - Heart Function A First Extension - CardiovascularBlood Flow 3D MRI Movies ?! - An (Almost) Unbelievable Story … Biomed NMR JF11/59

  15. Real-Time MRI of the Heart – Turbulent Flow Reconstructions Using Nonlinear Inversion 18 ms 18 ms Four-Chamber View Two-Chamber View RF-Spoiled Radial FLASH, TR/TE = 2.0/1.3 ms, 8o, 2 × 2 × 8 mm3, FOV 256 × 256 mm2 9 Spokes, 128 × 128 Matrix, Acquisition Time 18 ms, 55 Frames per Second S Zhang et al, J CardiovascMagnReson 12:39, 2010 Biomed NMR JF11/57

  16. Velocity-Encoded Phase-Contrast MRI in Real Time Reconstructions Using Nonlinear Inversion Two Acquisitions Using Highly Undersampled Radial FLASH (interleaved or sequential) with Two Bipolar Phase-Contrast Gradients Gvencof Opposite Polarity (through-plane flow) AA Joseph et al, NMR Biomed, 2011 Biomed NMR JF11/41

  17. Velocity-Encoded Phase-Contrast MRI in Real Time Reconstructions Using Nonlinear Inversion Ascending Aorta, Through-Plane Flow, venc = 150 cm s-1 Anatomy Magnitude Image RF-Spoiled Radial FLASH, TR/TE = 3.44/2.76 ms, 10o, 1.8 × 1.8 × 6 mm3, FOV 256 × 256 mm2 2 × 7 Spokes, 144 × 144 Matrix, Total Acquisition Time 48 ms, 21 Frames per Second AA Joseph et al, NMR Biomed, 2011 Biomed NMR JF11/37

  18. Velocity-Encoded Phase-Contrast MRI in Real Time Reconstructions Using Nonlinear Inversion Ascending Aorta, Through-Plane Flow, venc = 150 cm s-1 Phase-Contrast Map Velocity Profile (20 Heartbeats) RF-Spoiled Radial FLASH, TR/TE = 3.44/2.76 ms, 10o, 1.8 × 1.8 × 6 mm3, FOV 256 × 256 mm2 2 × 7 Spokes, 144 × 144 Matrix, Total Acquisition Time 48 ms, 21 Frames per Second AA Joseph et al, NMR Biomed, 2011 Biomed NMR JF11/38

  19. Velocity-Encoded Phase-Contrast MRI in Real Time Reconstructions Using Nonlinear Inversion Ascending Aorta, Through-Plane Flow, venc = 150 cm s-1 Dynamic Velocities During Single Heartbeat (12 Frames =512 ms) RF-Spoiled Radial FLASH, TR/TE = 3.05/2.29 ms, 10o, 1.8 × 1.8 × 6 mm3, FOV 256 × 256 mm2 2 × 7 Spokes, 144 × 144 Matrix, Total Acquisition Time 42.5 ms, 23 Frames per Second AA Joseph et al, NMR Biomed, 2011 Biomed NMR JF11/76

  20. Real-Time MRI – Outline Technical Considerations - Data Acquisition - Image Reconstruction Preliminary Applications - Joint Movements, Swallowing, Speaking - Heart Function A First Extension - CardiovascularBlood Flow 3D MRI Movies ?! - An (Almost) Unbelievable Story … Biomed NMR JF11/59

  21. Spatially Encoded Phase-Contrast MRI 3D MRI of 2D Objects Using 2 Projection Images Two Acquisitions of a Projection Image (nonselective RF excitation) with Two Monopolar Phase-Contrast Gradients Gpcof Opposite Polarity (perpendicular encoding) KD Merboldt et al, MagnReson Med 66:950-957, 2011 Biomed NMR JF11/25

  22. Spatially Encoded Phase-Contrast MRI 3D MRI of 2D Objects Using Two Projection Images Three-Dimensional Visualization of the 2D Phase-Contrast Projection Map 2D Phase-Contrast Projection Map Using Radial FLASH (color = phase difference) KD Merboldt et al, MagnReson Med 66:950-957, 2011 Biomed NMR JF11/26

  23. Spatially Encoded Phase-Contrast MRI in Real Time 3D MRI of the Acting Human Hand – 40 ms Resolution Reconstructions Using Nonlinear Inversion and Phase Difference RF-Spoiled Radial FLASH, TR/TE = 2.88/2.25 ms, 5o (nonselective), 2 × 2 mm2, FOV 256 × 256 mm2 2 × 7 Spokes (FOV 256 mm), 128 × 128 Matrix, Total Acquisition Time 40 ms, 25 Frames per Second KD Merboldt et al, MagnReson Med 66:950-957, 2011 Biomed NMR JF11/47

  24. Spatially Encoded Phase-Contrast MRI in Real Time 3D MRI of 1D Objects Using 3 MRI Signals Reconstructions Using Fourier Transformation and Phase Difference Three Acquisitions of a Frequency-Encoded MRI Signal (nonselective RF excitation) Two Orthogonal Monopolar Phase-Contrast Gradients Gpc1 and Gpc2 (two phase-encodings) KD Merboldt et al, MagnReson Med 66:950-957, 2011 Biomed NMR JF11/27

  25. Spatially Encoded Phase-Contrast MRI in Real Time 3D MRI of a Rotating Water-Filled Tube – 5 ms Resolution Reconstructions Using Fourier Transformation and Phase Difference Frequency-Encoded Gradient Echoes, TR/TE = 1.68/1.15 ms, 12o (nonselective), 2 × 2 mm2 3 Gradient Echoes, 128 × 128 Matrix, Total Acquisition Time 5 ms, 200 Frames per Second KD Merboldt et al, MagnReson Med 66:950-957, 2011 Biomed NMR JF11/22

  26. Real-Time MRI – Summary Data Acquisition – Low-Flip Angle Gradient Echo MRI (FLASH) - Radial Encoding with Pronounced Undersampling - Multi-Coil Data for Parallel MRI Image Reconstruction – Regularized Nonlinear Inversion - Joint Estimation of Image Content and Coil Sensitivities - Regularization with Previous Frame - Temporal and Spatial Filtering Current Projects - Reduction of Reconstruction Time (GPU Implementation) - Adaptation of Post-Processing Software (Cardiovascular MR) - Improvement of the Algorithm (Regularization, Filtering) - Many More Applications and Extensions Biomed NMR JF11/58