Optronic laboratories inc ol series 750 automated spectroradiometric measurement system
Download
1 / 28

Optronic Laboratories Inc. OL Series 750 Automated Spectroradiometric Measurement System - PowerPoint PPT Presentation


  • 444 Views
  • Uploaded on

Optronic Laboratories Inc. OL Series 750 Automated Spectroradiometric Measurement System . An Advanced Turnkey System for Automated Detector Response Measurements. - PMTs - Indium Antimodide (InSb) - Silicon (Si) - Mercury Cadmium Telluride (HgCdTe) - Germanium (Ge) - Pyroelectric

loader
I am the owner, or an agent authorized to act on behalf of the owner, of the copyrighted work described.
capcha
Download Presentation

PowerPoint Slideshow about 'Optronic Laboratories Inc. OL Series 750 Automated Spectroradiometric Measurement System' - emmett


An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -
Presentation Transcript
Optronic laboratories inc ol series 750 automated spectroradiometric measurement system l.jpg

Optronic Laboratories Inc. OL Series 750 Automated Spectroradiometric Measurement System

An Advanced Turnkey System for Automated Detector Response Measurements

- PMTs - Indium Antimodide (InSb)

- Silicon (Si) - Mercury Cadmium Telluride (HgCdTe)

- Germanium (Ge) - Pyroelectric

- Indium Gallium Arsenide (InGaAs) - CCD

- Lead Sulphide (PbS) - Cameras

- Lead Selenide (PbSe) - Radiometers and Photometers


About optronic laboratories inc l.jpg
About Optronic Laboratories, Inc.

Optronic Laboratories, Inc. was established in 1970 by two eminent researchers at NIST (then NBS). Facilities were modeled after NIST, and Optronic Laboratories, Inc. achieved a worldwide reputation for excellence in light measurement and calibration.

Continuing to work closely with NIST, the development of new instruments and techniques at Optronic Laboratories, Inc. constantly improves the accuracy and precision of light measurement.


About optronic laboratories inc3 l.jpg
About Optronic Laboratories, Inc.

In 1998, Optronic Laboratories, Inc. moved into its new 25,000 square foot purpose-built facility, consolidating and expanding production, R & D and calibration facilities


About optronic laboratories inc4 l.jpg
About Optronic Laboratories, Inc.

  • Spectroradiometers:

    • laboratory & field portable

Optronic Laboratories’ instruments & services include:

  • Broad-band radiometers & photometers:

    • DC, AC, pulsed or modulated sources

  • Sphere systems:

    • from 1” to 76” diameter as standard

  • NIST-traceable calibration sources, materials & detectors:

    • from 200 nm to 30,000 nm

  • Measurement services:

    • sources, materials & detectors

  • Customization/new product development

  • Training, consultation & advice


The ol series 750 monochromators l.jpg
The OL Series 750 Monochromators

The heart of any spectroradiometric system is the monochromator, but…

  • It must give the right bandpasses – not too large and not too small.

  • It must give high spectral purity (low stray light) so only the wavelength of interest is detected.


The ol series 750 monochromators6 l.jpg
The OL Series 750 Monochromators

  • Available in single and double (shown) versions

  • F/4 optics for high throughput

  • Very low stray light

  • Numerous special features for most accurate results


The ol series 750 monochromators7 l.jpg
The OL Series 750 Monochromators

  • Large range of gratings available

  • Optimised for high efficiency

  • Optimised for wide wavelength range

  • Optimised for high resolution

Tri-grating mounts change gratings automatically


The ol series 750 monochromators8 l.jpg
The OL Series 750 Monochromators

  • Range of Plug-in input accessories

  • Choppers (shown) for IR detectors

  • Automatic control of chopper speed and “open” position stop

  • Filter-wheel for input attenuation or filtering

Versatile plug-in chopper or filter wheel input accessories


The ol series 750 monochromators9 l.jpg
The OL Series 750 Monochromators

  • Fully automated 11 position blocking-filter wheel

  • Includes SHUT position for dark current readings

  • Optimised filters for widest wavelength range

Blocking-filter wheel changes filters automatically


The ol series 750 monochromators10 l.jpg

400nm

400nm

The OL Series 750 Monochromators

Blocking filters:

how they work

  • Gratings diffract light at different angles depending on wavelength

  • 400nm will also diffract at the same angle as 800nm (since 2 x 400 = 800)

  • A blocking filter removes the 400nm so just 800nm is left

Blocking filter

400nm + 800nm

800nm

grating


The ol series 750 monochromators11 l.jpg
The OL Series 750 Monochromators

Light passes into the monochromator through the entrance slit, then onto the first collimating mirror.


The ol series 750 monochromators12 l.jpg
The OL Series 750 Monochromators

Collimated (parallel) light is directed onto the grating.


The ol series 750 monochromators13 l.jpg
The OL Series 750 Monochromators

The grating diffracts the light, sending each wavelength off at a different angle.

The desired wavelength is collected by the first focussing mirror….


The ol series 750 monochromators14 l.jpg
The OL Series 750 Monochromators

…which focuses it at the middle slit.

A second collimating mirror collects this beam…


The ol series 750 monochromators15 l.jpg
The OL Series 750 Monochromators

…and collimates it onto a second grating.

Since the second grating is linked to the first, they are at the same angle and hence the same wavelength.


The ol series 750 monochromators16 l.jpg
The OL Series 750 Monochromators

This second grating acts like the first, but the spectral purity is increased by many decades.


The ol series 750 monochromators17 l.jpg
The OL Series 750 Monochromators

The pure monochromatic light, at the wavelength desired, is then focussed onto the exit slit and emerges from the monochromator.


The ol series 740 20 light sources l.jpg
The OL Series 740-20 Light Sources

  • Single and double lamp versions

  • UV lamp

    • 200nm to 400nm

  • Tungsten lamp

    • 250nm to 2500nm

  • Glowbar

    • 1000nm to 30000nm


The ol series 740 20 light sources19 l.jpg
The OL Series 740-20 Light Sources

  • Optics match the OL Series 750 Monochromators

  • Easy switching from one source …


The ol series 740 20 light sources20 l.jpg
The OL Series 740-20 Light Sources

…to the other

  • Optics match the OL series 750 monochromators

  • Easy switching from one source …

  • Fully automated switching available


The ol 750 ol 740 20 l.jpg
The OL 750 + OL 740-20

  • 200 to 30,000 nm coverage

  • Monochromatic light prevents sample heating

  • Suits a very wide range of standard OL accessories

  • Suits many existing and future applications


Detector spectral response l.jpg
Detector Spectral Response

  • In order to characterize their performance in a particular applications, single detectors, array detectors and assemblies often need spectral response to be known.

  • For example, photopic detectors can be analyzed for V() matching.


Detector spectral response23 l.jpg
Detector Spectral Response

  • Measurements are generally consistent with detector use and may require a large collimated beam, a small spot, a diffuse source or other geometry.

  • Optronic Laboratories, Inc. manufactures a wide range of accessories for most detector measurement requirements, including quantum efficiency.


Detector spectral response24 l.jpg
Detector Spectral Response

  • Optronic Laboratories, Inc. provides standard detectors, traceable to NIST and other National Laboratories.

  • To illustrate a detector spectral response system, we will use the OL 750-11C All Mirror Collimator.


The ol 750 11c l.jpg
The OL 750-11C

  • 1 inch diameter beam

  • Excellent beam uniformity

  • Constant collimation at all wavelengths


The ol 750 ol 740 20 ol 750 11c l.jpg
The OL 750 + OL 740-20 + OL 750-11C

If we put these together…


The ol 750 ol 740 20 ol 750 11c27 l.jpg
The OL 750 + OL 740-20 + OL 750-11C

We have a complete system for detector response

First we calibrate with the standard detector

Then we measure the test detector


More information l.jpg
More Information

  • The system illustrated represents one of the applications of the OL 750 Series ASMS, others include:

  • Source Spectral Analysis (Irradiance and Radiance)

  • Transmission (Diffuse and Linear)

  • Reflectance (Diffuse and Specular)

  • Visit our website at www.olinet.com