A Robot for Automated Motion Scaling in Robotic Microsurgery

A Robot for Automated Motion Scaling in Robotic Microsurgery. Group 2 Keshav Chintamani Lavie Golenberg Prashanth Mathihalli. Problem Statement. Different surgical tasks require varying motion scales (MS) between the surgeon and the end-effector

dorit
Download Presentation

A Robot for Automated Motion Scaling in Robotic Microsurgery

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. A Robot for Automated Motion Scaling in Robotic Microsurgery Group 2 Keshav Chintamani Lavie Golenberg Prashanth Mathihalli

  2. Problem Statement • Different surgical tasks require varying motion scales (MS) between the surgeon and the end-effector • E.g. Suturing vs. gross translation have different scale requirements • Surgeons currently keep the MS constant due to • Inconvenience and interruptions during procedures • Higher mental workload • Failures in selecting correct scales might lead to fatal errors • May require an additional technician

  3. Currently… • A touch screen allows the surgeon or technician to change the motion scale value • The scale is a semi circle with a minimum value of 1 and a maximum of 10 • Tapping the circumference changes the MS value of the Zeus robot

  4. Specific Aims • Design and construct an Automated Motion Scaling Robot (AMSR) • Hardware • Software • Integrate the AMSR with zeus robot system • Analysis and validation of the AMSR through an objective human factors study

  5. The Hypotheses The AMSR will: • Eliminate the need for a technician • Remove pauses during operations • Be capable of changing the MS more frequently • Deliver MS changes more accurately than a human • Not be susceptible to fatigue • Create a more responsive system

  6. Hardware

  7. Design Requirements • A two degree-of-freedom (DOF) RP robot arm • One rotational and one prismatic joint • Damping mechanism to prevent damage to the touch screen • Rapid input/output response • Provide accurate responses to inputs from the surgeon • Provide ease of removal during maintenance and repair

  8. Design Requirements, Contd.. • Designing the Robot Mount to provide • A high center-center accuracy between the AMSR and the Motion Scale • Variable chassis geometry settings for the AMSR for calibration

  9. Design Hurdles • Providing motors with sufficient torque • Providing a unique design that is • Replaceable • Reliable • Sensitive to pressure

  10. Preliminary Concepts for the Prismatic Joint

  11. Motor Selection • A high torque motor was chosen for the base (rotation) • 300 deg/sec Angular Velocity • 11 Kg/cm Peak Torque • A light weight motor for tapping (translational) • 24 g net weight • 3 Kg/cm Peak Torque • 350 deg/sec angular velocity

  12. Software

  13. Electronics • Robix RCS-6 Controller • Controller provides support for • 6 servos with 6 sensor inputs • Parallel port data transmission • The programming was done in Microsoft Visual C++ 6.0

  14. Robot Control Software • Fully integrated control functions for • Speed, acceleration and deceleration of servos • Positional feedback • Additional sensor data acquisition capabilities

  15. Final Design

  16. Final Design: 2D Views

  17. Final Design: 3D View

  18. A Descriptive Video

  19. The AMSR!

  20. Evaluation

  21. Methodology • Obtain preliminary data for 3 humans and the AMS Robot performing a tapping task • Compare performance between the robot and the subjects

  22. Preliminary Human Factors Test • Participants were provided with 5 minutes for practice on the MS display • They were asked to input 99 values based on verbal prompts from the experimenter • Participants were asked to tap values with and without a stylus • Values displayed on screen were recorded

  23. Analysis • A within-subjects factorial design was used • The experiment was balanced using a Latin square • The data was analyzed for input error

  24. Results, Discussion & Conclusion

  25. Plots Error variation between AMSR and Human Error variation between factors

  26. Overall Plots

  27. Conclusion • Human beings are incapable of the level of dexterity that robots possess • Hand movements with a stylus improved human performance • The AMSR can provide more rapid and accurate cyclic responses than a human • These responses are repeatable

  28. Conclusion, contd… • With the AMSR, surgeon performance can immensely be enhanced • Surgeon fatigue and workload can be reduced • Can result in efficient surgeries with reduced time durations

  29. Future Work

  30. Future Work • Creating a closed loop system • Increase the accuracy of the robot • Continue subject testing • Analyze performance of linear scales over semi-circular scales • Provide various forms of input methods

  31. Future Work, contd… • Combine AMSR with Automatic Motion Scaling • This study can lead to further research into human hand tracking performance • Develop display methods/cues for enhancing performance

  32. Thank you

  33. References • S. M. P. M. Sunil M. Prasad MD*, Hersh S. Maniar MD†, Celeste Chu MD*, Richard B. Schuessler PhD* and Ralph J. Damiano, Jr. MD*, Corresponding Author Contact Information, FACS, "Surgical robotics: Impact of motion scaling on task performance," 2004. • R. D. Ellis, A. Cao, A. Pandya, A. Composto, M. D. Klein, and G. Auner, "Minimizing Movement Time In Surgical Telerobotic Tasks," presented at 49th Annual meeting of the Human Factors and Ergonomics Society Orlando, Florida, 2005. • J. Accot and S. Zhai, "Scale effects in steering law tasks," CHI, vol. No.3, pp. 1-8, 2001. • P. M. Fitts, "The Information Capacity of the Human Motor System in Controlling the Amplitude of Movement," Journal of Experimental Psychology, vol. Vol. 121, pp. 262-269, 1954.