Resolution studies on a small animal RPC-PET prototype

Resolution studies on a small animal RPC-PET prototype. A. Blanco 1 , N. Carolino 1 , C.M.B.A. Correia 2 , L. Fazendeiro 1 , Nuno C. Ferreira 3 , M.F. Ferreira Marques 4,5 , R. Ferreira Marques 1,6 , P. Fonte 1,5 , C. Gil 4 , M. P. Macedo 2,5.

dkevin
Download Presentation

Resolution studies on a small animal RPC-PET prototype

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Resolution studies on a small animal RPC-PET prototype A. Blanco1, N. Carolino1, C.M.B.A. Correia2, L. Fazendeiro1, Nuno C. Ferreira3, M.F. Ferreira Marques4,5, R. Ferreira Marques1,6, P. Fonte1,5, C. Gil4, M. P. Macedo2,5 1 LIP, Laboratório de Instrumentação e Física Experimental de Partículas, 3004-516 Coimbra, Portugal. 2 CEI, Centro de Electrónica e Instrumentação, Univ. Coimbra, 3004-516 Coimbra, Portugal. 3 IBILI, Instituto Biomédico de Investigação de Luz e Imagem, Faculty of Medicine, 3000, Coimbra Portugal. 4 ICEMS, Departamento de Física, Universidade de Coimbra, 3004-516 Coimbra, Portugal. 5 ISEC, Instituto Superior de Engenharia de Coimbra, 3031-199 Coimbra, Portugal 6 Departamento de Física, Universidade de Coimbra, 3004-516 Coimbra, Portugal.

  2. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra • Summary • Small Animal PET overview. The basic idea of construction with RPCs. • The first prototype. • Experimental intrinsic and image spatial resolution. • Different contributions. Effect of positron range distribution. • Simulated performance • Conclusions Resolution studies on a small animal RPC-PET prototype

  3. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea HIDAC Micro PET-Focus YAP-PET Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra The small animal PET • Scintillator based technology (BGO, LSO, YAP, …), MWPC. • Ring diameter / axial length  150 mm / 50 mm => ~ low geometry acceptance • Spatial resolution  1 - 2 mm FWHM • Timing resolution ~ 2-3 ns • Central Point absolute sensitivity •  5 - 70 cps/KBq • Applications: Development of new drugs, human disease studies, validation of Gene therapies, ...

  4. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra The basic idea for a small animal PET with RPCs • High geometry acceptance > 90%. • Fully 3D measurement of the interaction point of the photon => No parallax error. • Sub-millimetre spatial resolution. • High timing resolution ~ 0.3 ns FWHM* • Moderate Efficiency. • Compatible with Magnetic Resonance Imaging. Use the plates as a g conversor * Nucl. Instr. And Meth A, 443 (2003) 88-93

  5. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra RPCs Charge-sensitive electronics allowing interstrip position interpolation 32 strips 16 plates 16 stacked RPCs. ....... ....... Aimed at verifying the concept and show the viability of a sub-millimetre spatial resolution. Depth of interaction Z 2D measurement of the photon interaction point. 3D measurement possible NIMA 508 (2003) 70–74 X Transaxial

  6. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea • Copper (on a PCB) and glass electrodes. • 32 1-mm wide X pickup strips. • 0.3 mm Gap. • Not optimized for high efficiency. Glass electrode (anode) Copper electrode (cathode) X Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra RPCs Transaxial coordinate Depth of interaction Active area 32 x 10 mm2 0.3 mm spacers

  7. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra Scanner

  8. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea 22Na source 0.22 Ø x 0.5 mm Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra Intrinsic spatial resolution: How to measure . LORs Red lines correspond to real data acquired with the 22Na source LOR = Line of Response. Connects the interaction points of the photons. D = Distance between each LOR and the center of the system

  9. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea • Annihilation photon non-collinearity. • N(x) = exp(-x2/2σ2), where FWHM = 2.35σ = 0.0022dswith ds the system diameter (mm). • Positron range. • P(x) = C1exp(-k1x) + (1-C1)exp(-k2x) • Source size. • S(x)=sqrt(size2 - x2), where size is the source radio (mm). • Detector response. • D(x) = exp(-x2/2σdet2) • Scatter background. • SC(x) = exp(-k3x) • R(x) = C2(N(X) P(X) D(X) S(X))+(1-C2)SC(X) • Annihilation photon non-collinearity. • N(x) = exp(-x2/2σ2), where FWHM = 2.35σ = 0.0022dswith ds the system diameter (mm). • Positron range. • P(x) = C1exp(-k1x) + (1-C1)exp(-k2x) • Source size. • S(x)=sqrt(size2 - x2), where size is the source radio (mm). • Detector response. • D(x) = exp(-x2/2σdet2) • Annihilation photon non-collinearity. • N(x) = exp(-x2/2σ2), where FWHM = 2.35σ = 0.0022dswith ds the system diameter (mm). • Positron range. • P(x) = C1exp(-k1x) + (1-C1)exp(-k2x) • Source size. • S(x)=sqrt(size2 - x2), where size is the source radio (mm). • Annihilation photon non-collinearity. • N(x) = exp(-x2/2σ2), where FWHM = 2.35σ = 0.0022dswith ds the system diameter (mm). • Positron range. • P(x) = C1exp(-k1x) + (1-C1)exp(-k2x) • Source size. • S(x)=sqrt(size2 - x2), where size is the source radio (mm). • Detector response. • D(x) = exp(-x2/2σdet2) • Scatter background. • SC(x) = exp(-k3x) • Annihilation photon non-collinearity. • N(x) = exp(-x2/2σ2), where FWHM = 2.35σ = 0.0022dswith ds the system diameter (mm). • Annihilation photon non-collinearity. • N(x) = exp(-x2/2σ2), where FWHM = 2.35σ = 0.0022dswith ds the system diameter (mm). • Positron range. • P(x) = C1exp(-k1x) + (1-C1)exp(-k2x) Phys. Med. Biol. 44 (1999) 781-799. 520 m FWHM 1550 m FWTM K2 = 3.75 mm-1 J. Nucl. Med. 34 101 1993. IEEE TNS, vol 33, No 1, (1986), 565-569 FWHMdet = 220 µm C2 = 0.04, K3 = 0.32 mm-1 Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra Intrinsic spatial resolution • Annihilation photon non-collinearity. • N(x) = exp(-x2/2σ2), where FWHM = 2.35σ = 0.0022dswith ds the system diameter (mm). • Positron range. • P(x) = C1exp(-k1x) + (1-C1)exp(-k2x) • Source size. • S(x)=sqrt(size2 - x2), where size is the source radio (mm). • Detector response. • D(x) = exp(-x2/2σdet2) • Scatter background. • SC(x) = exp(-k3x) Point spread function

  10. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra Intrinsic spatial resolution. Using a magnetic field Using an intense static homogeneous magnetic field => Improve spatial resolution ?? - Positron range is reduced(only two dimensions perpendicular to B) big effect with high energy radioisotopes. - Annihilation photon non-collinearity is reducedby mixing orthopositronium and parapositronium states,small effect. - Wirrwar et al. IEEE TNS Vol 44 No 2, April 1997. • Bibliography: • Hammer et al. Phy. Med. Biol. 40 (1995) 691-697. • Hammer et al. IEEE TNS vol 42 No 4, August 1995. • - Raylman et al. IEEE TNS vol 43 No 4, August 1996. Remember: This technology can operate in a magnetic field

  11. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra Intrinsic spatial resolution. Geant4 simulations with B • Comparing G4 with a measurement using 68Ga Geant4 * Positron Range  Adjusted Gaussian Response 0 T (1.0 mm FWHM, 5.7 mm FWTM) 5 T (1.1 mm FWHM, 2.8 mm FWTM) Geant4 => Geant4 compares reasonably well with this measurement.

  12. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra Intrinsic spatial resolution. Geant4 simulations with B Geant4 simulation for 22Na • The effect for 22Na is very small even at 10 T. • The idea is still valid for high energy radioisotopes (13N, 68Ga, 82Sr).

  13. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra Intrinsic spatial resolution: Limits Spatial resolution calculated as: R = C2*(N  P  D  S) + (1-C2)*SC Diameter system 60 mm RPC-PET 18F => 520m Physical limit 18F => 350 m positron range + non-collinearity

  14. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra Image spatial resolution. Positron range contribution The positron distributioncontributes in different ways to the sinogram and to the final image. 1D-histogram of the 3D-distribution contribution to: Sinogram PSF (intrinsic resolution) 2D-histogram profile of the 3D-distribution contribution to: Image PSF (image resolution)

  15. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea Proceeding IEEE MIC (2004) M2-177 Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra Image spatial resolution. Maximum likelihood-expectation maximization with resolution modeling (ML-EM) ~ 305 m FWHM Filtered Back Projection FBP ~ 465 m FWHM 0.46 0.31 0.47 0.48 0.45 0.30 0.32 0.29 Homogeneous spatial resolution over the entire detector

  16. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra Simulated count rate performance Evaluation of the count rate performance Prompts, Randoms, and NECR Mouse size phantom: 25 Ø x 70 mm (34 cm3) Annihilations and photons on the detector surface from the simulation. • Characteristic of the simulated system: • 90% Solid angle coverage => defining a FOV of 60 Ø x 100 mm axial. • Narrow coincidence window 1 ns. (Timing resolution 300 ps FWHM*) • Dead time ~ 100 ns. • 10% - 15% detection efficiency. * Nucl. Instr. And Meth A, 443 (2003) 88-93

  17. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra Simulated count rate performance: Prompts 1000 Kcps True + Scatter 500 Kcps Typical injected doses for mouse studies (3.7 MBq – 37 MBq)

  18. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra Simulated count rate performance: Randoms The high timing resolution keeps the random events at a low values Typical injected doses for mouse studies (3.7 MBq – 37 MBq)

  19. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra Simulated count rate performance: NEC 318 Kcps 240 Kcps Typical injected doses for mouse studies (3.7 MBq – 37 MBq)

  20. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra Performance Comparison between different small animal PET parameters and the expected parameters of the RPC-PET. • Yuan-Chuan Tai et al., “MicroPET II: design, development and initial performance of an improved MicroPET scanner for small-animal imaging”, Phys. Med. Biol. 48 (2003) 1519-1537. • Yongfeng Yang, et al., “Optimization and performance evaluation of the microPET II scanner for in vivo small-animal imaging”, Phys. Med. Biol. 49 (2004) 2527-2545. • A. del Guerra, G. Di Domenico, M. Scandola, G. Zavattini, “YAP-PET: first results of a small animal Positron Emission Tomograph based on YAP:Ce finger crystals”, IEEE Trans. Nucl. Sci., vol 45, No. 6 December 1998, 3105-3108. • G. Di Domenico et al., “Characterization of the Ferrera animal PET scanner”, Nucl. Instr. And Meth. A, 477 (2002) 505-508. • A.P. Jeavons, R.A. Chandler, C.A.R. Dettmar, “A 3D HIDAC-PET Camera with Sub-millimetre Resolution for Imaging Small Animals”, IEEE Trans. Nucl. Sci., vol. 46, No. 3, June 1999, 468-473. • Richard Laforest et al. “Performance Evaluation of the microPET – Focus F120”, presented at IEEE NSS/MIC Rome 2005.

  21. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra Conclusions • Results concerning the spatial resolution of an RPC-PET prototype have been presented. • An intrinsic spatial resolution of 520 mm FWHM with a intrinsic detector resolution of 220 mm FWHM have been measured. The resolution is largely dominated by intrinsic physical limits. • We studied by simulation the possibility to reduce the physical contributions using magnetic fields. Only for high energy radioisotopes. • Measured image spatial resolution of 305 mm FWHM using a ML-EM algorithm, homogeneous over the entire detector. • Asimulated peak NECR of 318 Kcps have been obtained for a optimized system dedicated to small animal PET. • This technology seems to be very appropriate for small animal PET studies, providing a very high spatial resolution and high sensitivity at a low cost.

  22. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea Groups of sums of 4 charges => Save electronic Q1 Q2 Q3 Q4 Interstrip position interpolation Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra

  23. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra

  24. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra

  25. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra Image spatial resolution: Construction Small range of angles available Data are transported to the rest of sinograma through the known sinogram curve

  26. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra Simulated count rate performance: NEC 318 Kcps 240 Kcps NECR - The number of true coincidences that would create an image of similar quality in the absence of noise (scattered and random coincidences). Typical injected doses for mouse studies (3.7 MBq – 37 MBq)

  27. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea Efficiency Vs nº stacked RPCs 0.25 All interactions 30 60 100 0.2 140 0.15 Detecting Efficiency 0.1 Counter-scattered 0.05 0 50 150 250 350 450 550 Energy (KeV) Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra Simulated count rate performance Mouse size phantom: 25 Ø x 70 mm (34 cm3) Geant 4 Annihilations and photons on the detector surface from the simulation. • Characteristic of the simulated system: • 90% Solid angle coverage => defining a FOV of 60 Ø x 100 mm axial. • Narrow coincidence window 1 ns. (Timing resolution 300 ps FWHM) • Dead time ~ 100 ns. • 10% - 15% detection efficiency.

  28. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra Simulated count rate performance Evaluation of the count rate performance Randoms, Prompts, Scatter and NECR Mouse size phantom: 25 Ø x 70 mm (34 cm3) NECR – Noise equivalent Count Rate The number of true coincidences that would create an image of similar quality in the absence of noise (scattered and random coincidences). Where T = True events, S = Scatter events, R = Random events Annihilations and photons on the detector surface from SIMSET. • Characteristic of the simulated system: • 90% Solid angle coverage => defining a FOV of 60 Ø x 100 mm axial. • Narrow coincidence window 1 ns. (Timing resolution 300 ps FWHM) • Dead time ~ 100 ns. • 10% - 15% detection efficiency.

  29. RPC2005 VIII Workshop on Resistive Plate Chambers. October 10-15, 2005. Seoul, Korea Resolution studies on a small animal RPC-PET prototypeAlberto Blanco. LIP-Coimbra RC passive network RC passive network X-strips (deposited on glass) Single-gap timing RPCs with bidimensional position-sensitive readout for very accurate TOF systems NIMA 508 (2003) 70–74 TESTED X right X left out left 10 strips for each coordinate at 4 mm pitch Y-strips (on PCB) out right