NCAR Airborne Oxygen Instrument (AO2). measures O 2 concentration (reported as O 2 /N 2 ratio) using a vacuum-ultraviolet absorption technique

Download Presentation

NCAR Airborne Oxygen Instrument (AO2)

An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -

Presentation Transcript


Ncar airborne oxygen instrument ao2

NCAR Airborne Oxygen Instrument (AO2)

  • measures O2 concentration (reported as O2/N2 ratio) using a vacuum-ultraviolet absorption technique

  • based on earlier shipboard and laboratory instruments, but designed specifically for airborne use to minimize motion and thermal sensitivity and with a pressure and flow controlled inlet system

  • flew on the Wyoming King-Air during ACME-2007

  • switches between sample gas and air from a high-pressure reference cylinder every 2 seconds and has a precision of +/- 5 per meg on a 5-second measurement

  • consists of a pump module, a cylinder module, an instrument module, and a dewar


Ncar airborne oxygen instrument ao2

NCAR Airborne Oxygen Instrument (AO2)

Contact name :Britton Stephens, NCAR EOL/TIIMES, 303-497-1018, [email protected] : 19" rack wide X 30" tall X 22" deep plus a 6" diameter X 14" tall dewar

Instrument weight :200 lbs.

Power:110 AC

Hazmat : 4 kg of solid CO2 (dry ice)

Inlet(s) : dedicated aft facing 1/4" inlet, unheated stainless steel or Synflex

secondary inlet to rule out fractionation

Number of operators :1 (eventually planned for 0 as instrument matures)

Altitude requirements :max. alt 60,000 feet

Ground requirements :30 minute preflight

[Stephens, B.B., R.F. Keeling, and W.J. Paplawsky, Shipboard measurements of atmospheric oxygen using a vacuum-ultraviolet absorption technique, Tellus Series B, 55, 857-878, 2003.]


Ncar airborne oxygen instrument ao2

NCAR MEDUSA Flask Sampler

  • collects cryogenically dried air into 16 1.5 L glass flasks under actively controlled flow and pressure conditions

  • designed to sample for laboratory measurements of O2/N2, Ar/N2, 13CO2, and C18OO

  • earlier version flew on the UND Citation II (COBRA-2000 and 2003) and on the NCAR C130 (IDEAS, ACME-2004)

  • integrated control and data acquisition system logs housekeeping variables every second

  • consists of a pump module, a flask module, and a dewar


Ncar airborne oxygen instrument ao2

NCAR MEDUSA Flask Sampler

Contact name : Britton Stephens, NCAR EOL/TIIMES, 303-497-1018, [email protected]

Dimensions : 19" rack wide X 27 " high X 18 " deep, plus a 6" diameter X 14" high dewar

Weight : 100 lbs.

Power : 28 VDC

Hazmat : 4 kg of solid CO2 (dry ice)

Inlet : dedicated forward facing 3/8" inlet, unheated stainless steel or Synflex

Number of operators : 1 (can be shared)

Altitude requirements : max. alt 52,000 feet

Ground requirements : 30 minute preflight

COBRA-NA 2003

[Graven et al., Causes of Observed Summertime Variability in Tropospheric O2/N2 above North America, manuscript in preparation, 2008]


Ncar airborne oxygen instrument ao2

NCAR MEDUSA Flask Sampler

  • nominal flow rate is 3.3 SLPM and the flask volume is 1.5 L, resulting in a 30-second quasi-integrated sample.

  • GV cabin throughflow ~ 24,000 SLPM (unofficial value fromGulfstream)

  • Flask rupture would lead to an addition of < 3 SLPM of air into cabin (~ + 0.01 % change)

  • Flask rupture combined with upstream flow-controller failure to full-closed and downstream flow-controller failure to full-open would lead to a removal of < 4 SLPM of air from cabin (~ - 0.015%)