1 / 34

Computed Tomography SCANCO User Meeting 2005

Computed Tomography SCANCO User Meeting 2005. Dr. Bruno Koller SCANCO Medical AG www.scanco.ch. Overview. X-Ray Basics CT Hardware Components Measurement Reconstruction Artefacts. Introduction. 3D distribution of tissue-properties Density (absorption of X-rays, speed of sound…)

cole-lynn
Download Presentation

Computed Tomography SCANCO User Meeting 2005

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Computed TomographySCANCO User Meeting 2005 Dr. Bruno Koller SCANCO Medical AG www.scanco.ch

  2. Overview • X-Ray Basics • CT Hardware Components • Measurement • Reconstruction • Artefacts

  3. Introduction • 3D distribution of tissue-properties • Density (absorption of X-rays, speed of sound…) • Chemical composition • Temperature • ... • Imaging of these local tissue properties using grayscale or color mapping

  4. Introduction

  5. Whole Body CT • Good S/N • Good contrast bone/soft tissue • Slice thickness 2-5 mm 2 cm

  6. Peripheral CT • Good Contrast Bone/Soft tissue • Voxelsize 100 mm • Limited FOV (130 mm) 1 cm

  7. Microtomography • Excellent contrast bone/soft tissue • Slice thickness and in plane resolution <10 mm • More noise in images 1 mm

  8. 3D Microtomography

  9. CT-Basics • Based on measurement of attenuation of X-rays (Beer-Lambert): • Measurement of a projection value (Sample): Source Detector m Io I d

  10. Measurement of one projection

  11. Measurement of one projection

  12. Measurement of one projection

  13. Measurement of one projection Io I t t

  14. Projection Value Measurement I I I0 m X-rays Source Object Detector

  15. Source • X-Ray Tubes (most common) • Continuous, steady output (high flux) • Small focal spot (< 10 mm) • Variable energy and intensity • Polychromatic beam I E

  16. Attenuation coefficient m [1/cm] • Attenuation coefficient changes with material: • Attenuation coefficient changes with energy: m bonemusclefat E

  17. I I E E Beam Hardening • Soft X-rays are attenuated more than hard X-rays • Depending on object, spectrum changes m(E) d

  18. Detectors • Usually detect visible light only • Counting Systems (Photomultipliers) • Integrating Systems (CCD, Diode Arrays, CMOS-Detectors) • They all need Scintillators • Convert X-rays into light • NaI, CsI, CdTe ... • The thicker, the more efficient, but the thiner, the better the spatial resolution (tradeoff between high output or high res) • Fiber optics (straight or tapered) in between to protect from remaining X-rays

  19. CT-Measurement • For a CT measurement one needs an certain number of single projection measurements at different angles (theoretically, an unlimited number is required) • In realized Tomography-Systems one usually finds a geometrically ordered detector configuration

  20. 1st generation scanner • Single Detector System • Translation-Rotation • 5 min. per slice

  21. 2nd generation scanner • multichannel-Systems (4, 6, 8, 16) • Translation-Rotation • 20 sec. per slice

  22. 3rd generation scanner • Fan-Beam-Geometry • multichannel-system (500+ detectors), angle > 180o • Rotation of tube and detektorsystem • no translation • 1 – 10 sec. per slice

  23. Parallel Beam (Synchrotron) • Parallelbeam • Rotation of object only • No collimators required • 2D-Detector arrays  A. Kohlbrenner, ETH Zürich

  24. Cone Beam • Tube with focal spot • Linear, 2-D Detector (e.g. 1024 x 1024 Elements, CCD) • Single rotation • Artefacts due to improper scanning scheme (would require to different movements)  A. Kohlbrenner, ETH Zürich

  25. Spiral scanning • Continuous movement of patient during rotation • Volumetric measurement • Slicewise reconstruction with variable slice thickness by interpolation • As scanner can continuously rotate, one can achieve much faster scan speeds • Latest models (clinical scanners) with parallel detector rings (Multirow, currently up to 64) • 40 slices per second (150 rpm) • No need in current MicroCT systems as the rotation speed is low

  26. Reconstruction • Iterative reconstruction • ART (Arithmetic Reconstruction Technique) • Assume image (base image) • Calculate projections of this base image • Modify image after comparing calculated projections with measured Projections • Strategy...

  27. Reconstruction • Direct method: The measured projections are backprojected under the same angle as the measurement was taken. All projections are summed up

  28. Reconstruction

  29. Reconstruction

  30. Convolution-Backprojection

  31. Artefacts • Beam Hardening • Attenuation coefficients depend on energy • soft X-rays are much more absorbed than harder X-rays • Distribution changes when beams penetrate object • Segmentation problems

  32. Artefacts • Object outside of FOV • Inconsistent set of projection data (only partially within the beam at some angles, completely in the beam at other angles) • Local Reconstruction: only for geometry

  33. Artefacts • Motion • Object moves during scan • May be eliminated by external gating (respiratory, heart beat) • Total absorption of X-Rays e.g. Caused by metallic implants (division by 0 in reconstruction) • Other Artefacts • Wrong geometry (fan-beam-angle) • Centers artefact • Mechanical alignment • Insufficient no. of projections (sampling) • ...

  34. Resources • Volume of 1024 x 1024 x 1200 requires 2.4 GB (short integer) • Doubling the resolution requiers 8x more time to calculate • Doubling the resolution requiers 8x more disk space

More Related