ABX Pentra 400
This presentation is the property of its rightful owner.
Sponsored Links
1 / 18

ABX Pentra 400 Specifications Reagents Main features analysis Strengths PowerPoint PPT Presentation


  • 101 Views
  • Uploaded on
  • Presentation posted in: General

ABX Pentra 400 Specifications Reagents Main features analysis Strengths.  Sample Tray.  Reagent Tray.  ISE Module. Physical Specifications.  Cuvette changer. . . . .  Windows XP. . Physical Specifications. Bench-top analyzer Dimensions : 100 x 65 x 57 cm

Download Presentation

ABX Pentra 400 Specifications Reagents Main features analysis Strengths

An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -

Presentation Transcript


  • ABX Pentra 400

  • Specifications

  • Reagents

  • Main features analysis

  • Strengths


 Sample Tray

 Reagent Tray

 ISE Module

Physical

Specifications

 Cuvette changer

 Windows XP


Physical

Specifications

Bench-top analyzer

Dimensions :

100 x 65 x 57 cm

Weight : 120 kg

Screen:

12 inches

57 cm

65 cm

100 cm


Technical

Specifications

Analysis Method Types

- Spectrophotometry : Colorimetry and Turbidimetry,

- Potentiometry : Direct (Serum or Plasma) and Indirect (Urine)

On-board Test Capacity : 55 parameters on board

- 52 cassette positions or 39 tests positions with Reagent racks

- 3 ISE's (Na+, K+, Cl-)


Technical

Specifications

  • External Modules

    • Cooling unit

  • Options

    • ISE Module

    • P400 Stand Alone

ISE Module


Technical

Specifications

Throughput :

- Without ISE : up to 300 Tests per hour

- With ISE : up to 420 Tests per hour


Technical

Specifications

  • Sample Handling

  • Sample Capacity

  • - 60 samples on board + Continuous loading

  • - 6 racks with 10 sample positions for primary tubes or cups

  • Positive Sample Identification

    Integrated Barcode Readerand tube presence detector

    Sample Needle

    - Level detection

    - Shock detection

    - Clot detection


    Technical

    Specifications

    • Sample Handling

    Sample Specimen

    Serum, Plasma, Urine, CSF, Whole blood, Homogeneous fluids

    Sample Volume

    - Photometry mode:2 to 95 µL (380 in 4 steps)

    - ISE mode : 65µL

    Automatic Sample Dilution

    from 1/2 to 1/22 500

    Automatic Post Concentration

    from x2.0 to x10.0


    Technical

    Specifications

    • Calibrator and Control Handling

    Calibrator/Controls Positioning

    Sample Tray or Reagent Tray

    Automatic Standard Dilution Series

    - Dilution ratio:main direct : 1/2 - 1/150

    - Dilution ratio: main indirect : 1/2 - 1/1508


    Technical

    Specifications

    • Reagent Handling

    Number of Reagents on-board : 52 positions

    - 44 in a closed refrigerated area (2-10°C)

    - 8 at room temperature

    Positive Reagent Identification

    Integrated barcode reader

    Reagent Needle

    - Capacitive Level detection

    - Preheating


    Technical

    Specifications

    • Reagent Handling

    Reagent Management with Cassettes

    - Automated Reagent On board stability check plus test countdown

    - Unique cassette barcode

    - back-up cassette

    Reagent Volume

    - Minimum : 2µL with Sample needle

    15µL with Reagent needle

    - Maximum : 600µL


    Technical

    Specifications

    • Reaction Handling

    Reaction Cuvettes

    - Disposable acrylic cuvettes

    - Cuvette volume : Min 150µl - Max 600µl

    Automatic Cuvette Changer

    - Continuous loading and unloadind of cuvette segments

    - Total Capacity : 36 segments (432 cuvettes)


    Technical

    Specifications

    • Reaction Handling

    Reaction Temperature

    Temperature controlled Air bath at 37°C

    Mixing

    By Rotation

    Incubation Time

    1 to 100 cycles (12 seconds to 20 minutes)


    Specifications

    Reagents

    Cassette Features

    • - 4 types of cassette (1 single and 3 twin)

    • - Side barcode indentifies for :

    • - type of reagent

    • - expiration date

    • - filling volume

    • - index of the cassette

    • - scrambling key

    • - Cassette Packaging

    • - On board Reagent Stability


    Specifications

    Reagents

    Advantages of the cassette

    • - Easy to handle

    • - Automatic Reagent management

    • - Lot number

    • - Open stability handled

    • - Test countdown

    • - Back-up cassettes


    ABX Pentra Range

    Characteristics

    • 52 reagents (30 in cassette format)

      • 9 Enzymes

      • 21 Substrates

      • - 22 Specific Proteins


    Compact Benchtop

    Integrated System

    52 refrigerated positions

    Comprehensive menu + Open channels

    Bar coded RTU Cassette

    3 ISE’s : Na, K, Cl

    High Autonomy

    Optical system

    Holographic grating technology

    Choice of 15 wavelengths from 340-700 nm

    Safe and secure environment

    Sampling constant check

    GLP/Logs

    Reagent management

    Appealing Easy to use software

    Integrated Validation Station

    Help on line

    Windows XP touch screen

    Main Features analysis


    Strengths

    • High autonomy

      Throughput

      Cuvette capacity

      Sample and reagent on-board capacity

    • Comprehensive menu

      Routine and Proteins

    • Reagent management

      RTU cassettes, easy handling

      Higher reagent stability

      Test countdown, lot and shelf life

      Cassette / vial flexibility

    • Easy to use and Intuitive software

      Touch screen interface

      Integrated validation station

      Remote diagnostic capabilities

      Contextual help

    • Flexibility

      • Closed system with number of

      • open channels # configurable

      • Cal / Ctrl placement on system

      • Most liquid chemistries adaptable


  • Login