СС E measurements with alpha source (241Am) on the test structures

СС E measurements with alpha source (241Am) on the test structures. Strip-pixel test structure: 7x7 mm^2; pitch=15 mkm; Y=8 strips; X=7 strips. Set up: - all 8 Y-strip and 7 X-strip shorted US-bonding and connected with two pins (Y-pin and X-pin);

chelsa
Download Presentation

СС E measurements with alpha source (241Am) on the test structures

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. ССE measurements with alpha source (241Am) on the test structures • Strip-pixel test structure: 7x7 mm^2; pitch=15 mkm; Y=8 strips; X=7 strips. • Set up: - all 8 Y-strip and 7 X-strip shorted US-bonding and connected with two pins (Y-pin and X-pin); - n+ Ohmic contact glued with Ag-glue on the Au-layer ceramics PCB; - CS-Amplifier connected with: 1. Y+X; 2. Y; 3. X. N.Zamyatin, JINR, 20-21 Oct., Prague

  2. Test structures: two strip-pixel 15mkm (#1-1; #2-1) and two PADs N.Zamyatin, JINR, 20-21 Oct., Prague

  3. Alpha spectrum (241Am) for test-PAD at bias 50 V, alpha source without collimator (putted on the p+ contact) N.Zamyatin, JINR, 20-21 Oct., Prague

  4. Alpha spectrum (241Am) for test#2-1_strip-pixel at bias 50 V, all Y- and X- strips connected together, alpha source without collimator (putted on the p+ contact). Max spectra=2912 ch, it’s correspond 100% CCE N.Zamyatin, JINR, 20-21 Oct., Prague

  5. Y-alpha spectrum (241Am) for all 8 Y-strips test#2-1 at bias 50 V, all 7 X-strips connected with the equal Y-strips RC-value (C=Cdinamic/PAmp), alpha source without collimator (putted on the p+ contact.- Edge of the max Y-spectra shifted on the left from 100%, because X-part of the signals collected on X-strips.- This spectra indicated, that all time should be signals (on Y- or on X- strips) with small amplitudes, that is problem of Signal/Noise!!! N.Zamyatin, JINR, 20-21 Oct., Prague

  6. X-alpha spectrum (241Am) for all 7 X-strips test#2-1 at bias 50 V, all 8 X-strips connected with the equal X-strips RC-value (C=Cdinamic/PAmp), alpha source without collimator (putted on the p+ contact.- Edge of the max X-spectra shifted on the left, because Y-part of the signals collected on Y-strips.- Note: X-strips more noisly N.Zamyatin, JINR, 20-21 Oct., Prague

  7. Y-alpha spectrum (241Am) for all 8 Y-strips test#2-1 at bias 50 V, all 7 X-strips are “floating”, alpha source without collimator (putted on the p+ contact.- Edge of the max Y-spectra have two pick, this effect connected with floating X-strips and interstrip capacitance “cross talk Y-X” N.Zamyatin, JINR, 20-21 Oct., Prague

  8. Conclusions : • Measurements of the alpha spectra for the strip-pixel sensors is very easy and informative method for the study of the Charge Collection on X- and Y- strips; • Preliminary results on the 15 mkm Strip-Pixel structures indicated about big amplitudes range of the Y- and X-signals (signal Y and X not 50% from total alpha pick). For the events with big Y(or X)-amplitudes we have small X(or Y)-signals and correspond low S/N, that is problem of the 100% efficiency on both Y- and X- coordinate; • For the ideal topology of the Strip-Pixel sensors should be the ratio Qy or Qx/Qy+Qx=0.5; • What to do?: - increase thickness up to 0.6mm; decrease pitch (?) or ZIP-shape (?); come back to SS-strip sensors. N.Zamyatin, JINR, 20-21 Oct., Prague