ELEVATION  AND PACHYMETRY VALUES IN NORMAL CORNEAS OBTAINED BY GALILEI
Download
1 / 12

ELEVATION AND PACHYMETRY VALUES IN NORMAL CORNEAS OBTAINED BY GALILEI - PowerPoint PPT Presentation


  • 154 Views
  • Uploaded on

ELEVATION AND PACHYMETRY VALUES IN NORMAL CORNEAS OBTAINED BY GALILEI. Claudia Blanco Marín MD. Maria Ximena Núñez MD. Vision Sana Research Group Cornea and Refractive Surgery Unit Clinica de Oftalmologia de Cali Cali-Colombia. ASCRS Boston April 9 -14 2010.

loader
I am the owner, or an agent authorized to act on behalf of the owner, of the copyrighted work described.
capcha
Download Presentation

PowerPoint Slideshow about ' ELEVATION AND PACHYMETRY VALUES IN NORMAL CORNEAS OBTAINED BY GALILEI' - channer


An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -
Presentation Transcript

ELEVATION AND PACHYMETRY VALUES IN NORMAL CORNEAS OBTAINED BY GALILEI

Claudia Blanco Marín MD.

Maria Ximena Núñez MD.

Vision Sana Research Group

Cornea and Refractive Surgery Unit

Clinica de Oftalmologia de Cali

Cali-Colombia

ASCRS Boston April 9 -14 2010

Theauthorshave no financialinterest in thesubjectmatter of this poster.


PURPOSE BY GALILEI

To obtain the topographic parameters in normal corneas of candidates for refractive surgery with the Galilei dual-Scheimpflug analyzerand compare them with the Orbscan II scanning-slit system.


METHODS BY GALILEI

  • Prospective study

  • Two measurements per eye from each patient and each device were taken by the same operator.

  • To obtain equal conditions in elevation measurements, the Orbscan II best fit sphere (BFS ) was obtained from an 8-mm-diameter area like the Galilei system does.

  • This study was analyzed with descriptive statistics and an univariate analysis of variables (Galilei and Orbscan) was performed in order to calculate:

    - The central tendency (mean , median) and confidence interval (CI)

    -The distribution (normal distribution test: Kolmogorov-Smirnov test)

    -The dispersion (Standard Deviation, SD)

  • To evaluate the differences in means between the two groups of gathered data (Galilei and Orbscan II) the t-test (parametric test) was used , with a level of significance of p<0.005.


SUBJECTS BY GALILEI

  • 92 eyes were examined with the two devices.

  • Gender : 25 males / 67 females

  • Ages 17 to 55 years old, mean: 32,77 years old (95% CI 30,57 - 34,96)

  • DX: Myopia 23% Myopic astigmatism 53,2%

    Hyperopia11% Hyperopic astigmatism 7,5%

    Presbyopia3,3 % Mixed astigmatism 2 %

  • We excluded keratoconus and any keratoconus suspects identified with the CLMI index from the Keratron Scout and the KPI from the Galilei .


Results anterior best fit sphere radii rabfs
RESULTS BY GALILEIAnterior Best-fit Sphere Radii (raBFS)

The mean radius of the anterior best-fit sphere (raBFS) was:

7.84 +/- 0.23 mm (SD)

(95% CI 7,795 -7,891)

with the Galilei

7.84 +/- 0.23 mm (SD)

(95% CI 7,791 -7,890)

with the Orbscan II

p= 1,0


Results posterior best fit sphere radii rpbfs
RESULTS BY GALILEIPosterior Best-fit Sphere Radii (rpBFS)

The mean radius of the posterior best-fit sphere (rpBFS) was:

6.50 +/- 0.20 mm(SD) (95% CI 6,46 -6,54)

with the Galilei

6.46 +/- 0.20 mm (SD) (95% CI 6,41 -6,50)

with the Orbscan II

p=0,17.


Results radius of the maximum anterior elevation rmae
RESULTS BY GALILEIRadius of the Maximum Anterior Elevation (rMAE)

Sample size =92

Lowest value =1,5

Highest value = 12

Mean = 5,61

SD = 2,19

Sample size = 92

Lowest value = 2

Highest value = 14

Mean = 4,64

SD = 2,23

The mean radius of the maximum anterior elevation (rMEA) was :

5,61 +/- 2,19 µm(SD)

(95% CI 5,16 – 6,07 )

with the Galilei

4,64 +/- 2,23µm (SD),

(95% CI 4,18 – 5,11)

with the Orbscan II

p= 0,0023


RESULTS BY GALILEIRadius of the Maximum Posterior Elevation (rMPE)

Sample size = 92

Lowest value = 5,50

Highest value = 22,00

Mean = 13,00

SD = 3,35

  • The mean radius of the maximum posterior elevation (rMPE) was:

  • 13,0 +/- 3,35 µm (SD)

  • (95% CI 12,31 – 13,69)

  • with the Galilei

  • 13,84 +/- 6,65 µm (SD),

  • (95% CI 12,47 – 15,22)

  • with the Orbscan II

  • p= 0,321

Sample size

= 92

Lowest value

= 1,00

Highest value

= 32,00

Mean = 13,84

SD= 6,65


Results central corneal thickness cct
RESULTS BY GALILEICentral Corneal Thickness (CCT)

  • The Central Corneal Thickness (CCT) mean was:

  • 549 +/- 29.96 microns (SD)

  • (95% CI 542 – 555)

  • with the Galilei,

  • Lowest 474 highest 612

  • 551 +/- 38.52 microns(SD)

  • (95% CI 543 – 559)

  • with the OrbscanII

  • Lowest 459 highest 644

  • p=0,722


Results corneal thinnest point ctp
RESULTS BY GALILEI Corneal Thinnest Point CTP

Corneal Thinnest Point mean CTP was :

546 +/- 29.6 microns(SD)

(95% CI 540 – 552)

with the Galilei

Lowest 472 highest 611

543 +/- 37.8 microns (SD)

(95% CI 535 –551)

with the Orbscan II

p=0,59.


Conclusion
Conclusion: BY GALILEI

There were a statistical difference between the two devices in the maximun anterior elevation with a better normal tendency of the measurements in the Galilei results.

The pachymetry analysis shows a higher dispersion of the data for the Orbscan II ; due to a SD of 38.52 microns(µ) for the Orbscan II and 29.96 µ for the Galilei in the CCT and 37.8 µ for the Orbscan II and 29.6 µ for the Galilei in the CTP.


Conclusion1
Conclusion: BY GALILEI

Even though both devices have different centering systems, most of the elevation and pachymetry mean results did not show significant differences.

It is important to know the elevation and pachymetry variables in normal corneas, and their distribution to establish the clinical cut off with the Galilei dual-scheimpflug analyzer.