Application of UV/Vis Spectroscopy
This presentation is the property of its rightful owner.
Sponsored Links
1 / 38

Application of UV/Vis Spectroscopy Common Problems : a) Mixtures : PowerPoint PPT Presentation


  • 35 Views
  • Uploaded on
  • Presentation posted in: General

Application of UV/Vis Spectroscopy Common Problems : a) Mixtures : - blank samples often contain multiple absorbing species. - the absorbance is the sum of all the individual absorbencies A= A 1 + A 2 +A 3 + … = e 1 bc + e 2 bc + e 3 bc …

Download Presentation

Application of UV/Vis Spectroscopy Common Problems : a) Mixtures :

An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -

Presentation Transcript


Application of uv vis spectroscopy common problems a mixtures

Application of UV/Vis Spectroscopy

Common Problems:

a) Mixtures:

- blank samples often contain multiple absorbing species.

- the absorbance is the sum of all the individual absorbencies

A= A1 + A2 +A3 + … = e1bc + e2bc + e3bc …

- substances in both the blank and sample which absorb can be “blanked out” in both double and single beam spectrometers.


Application of uv vis spectroscopy common problems a mixtures

But, If the blank absorbance is high, Po will decrease too much, the response will be slow and the results inaccurate

If blank absorbance too high:

- dilute the sample

- use a different l where the analyte absorbs more relative to the interference.

- use a different method of separation

Large blank

absorbance

l scan of substance


Application of uv vis spectroscopy common problems a mixtures

b) Instrumental Deviations from Beer’s Law:

- stray light (already discussed).

- polychromatic light (more then a single l)

‚since all instruments have a finite band pass, a range of l’s are sent through the sample.

‚e may be different for each l

Deviation’s from Beer’s law at high concentration


Application of uv vis spectroscopy common problems a mixtures

Illustration of Deviation from Beer’s law:

Let us say that exactly 2 wavelengths of light were entering the sample

l = 254 nm e254 =10,000

l = 255 nme255 = 5,000

let Po = 1 at both l’s

What happens to the Beer’s law plot as c increases?

A =A254 + A255

A = log Po/P (total) = log (Po254 + Po255)/(P254 + P255)

At the individual l’s:

A254 = e254bc = log Po254/P254

10ebc = Po254/P254

P254 = Po/10ebc = Po10-ebc

For both together:

A = log (Po254 + Po255)/(Po25410-e254bc + Po25510-e255bc)


Application of uv vis spectroscopy common problems a mixtures

Since Po254 = Po255 =1:

A = 2.0/(Po25410-e254bc + Po25510-e255bc)

A = 2.0/(10-10,000x1.0xc + 10-5000x1.0xc)

The results are the same for more l’s of light. The situation is worse for greater differences in e’s (side of absorption peak, broad bandpass)

Always need to do calibration curve! Can not assume linearity outside the range of linearity curve!


Application of uv vis spectroscopy common problems a mixtures

b) Chemical Deviations from Beer’s Law:

- Molar absorptivity change in solutions more concentrated than 0.01M

‚due to molecular interactions

‚Beer’s law assumes species are independent

‚electrolytes may also cause this problem

‚e is also affected by the index of refraction

- association, dissociation, precipitation or reaction of analyte

‚c in Beer’s law is the concentration of the absorbing species

‚commonly use the analytical concentration – concentration of all forms of the species.

Ka

phenolphthalein:

H+ + In-

HIn

Red, l =600nm

colorless

If solution is buffered, then pH is constant and [HIn] is related to absorbance.


Application of uv vis spectroscopy common problems a mixtures

But, if unbuffered solution, equilibrium will shift depending on total analyte concentration

Ka

example: if Ka = 10-4

H+ + In-

HIn

“Apparent” deviation since can be accounted for by chemical equilibrium


Application of uv vis spectroscopy common problems a mixtures

But, if unbuffered solution, equilibrium will shift depending on total analyte concentration

Ka

example: if Ka = 10-4

H+ + In-

HIn

Isosbestic point

At the isosbestic point in spectra:

A = eb([HIn] + [In-])


Application of uv vis spectroscopy common problems a mixtures

c) Non-constant b:

- worse for round cuvettes

- use parallel cuvettes to help

B2

P2

P0

B1

P1

A = log10 Po/P = ebc


Application of uv vis spectroscopy common problems a mixtures

d) Instrument noise:

noise - short term baseline fluctuations, which decrease the precision of the analysis

‚can not measure A precisely

‚the various sources of noise each cause some uncertainty

in the absorbance measurement and can be treated as individual standard deviations.

1) 0%T noise:

- noise when light beam is blocked

- seldom important

- typically ±0.01%T

2) Readout Precision:

- especially with a meter

- typically ±0.5%T  1-3% error in concentration

3) Shot Noise:

- occurs when e- transfers across a junction (like the space between cathode & anode in PMT).

- causes random fluctuations in current since individual e- arrive at random times

- increases with increase current (%T). Especially bad above 95%T.


Application of uv vis spectroscopy common problems a mixtures

4) Flicker Noise:

- noise from the lamp due to intensity changes

- important at high transmittances.

5) Cell positioning uncertainty:

- not really noise, but affects precision

- minor imperfections, scratches or dirt change %T

- may be the major cause of imprecision


Application of uv vis spectroscopy common problems a mixtures

Taken together, these noise sources

indicate that the intermediate absorbance

and transmittance ranges should be used.

- at low %T, 0%T, noise and readout precision

are important

- at high %T, shot and flicker noise are large.

Keep A in range of 0.1 – 1.5 absorbance units (80 -3%T)


Application of uv vis spectroscopy common problems a mixtures

Applications:

A) Molar Absorptivities (e) in UV-Vis Range:

e=8.7 x 1019 PA

P – transition probability (ranges from 0.1 to 1, for likely transitions)

A – cross-section area of target molecule (cm2)

- ~10-15 cm2 for typical organics

- emax = 104 to 105 L/mol-cm

- e < 103 – low intensity (P £ 0.01)


Application of uv vis spectroscopy common problems a mixtures

- For Compounds with Multiple Chromophores:

‚If greater then one single bond apart

- e are additive

- l constant

CH3CH2CH2CH=CH2lmax= 184 emax = ~10,000

CH2=CHCH2CH2CH=CH2lmax=185 emax = ~20,000

‚If conjugated

- shifts to higher l’s (red shift)

H2C=CHCH=CH2lmax=217 emax = ~21,000


Application of uv vis spectroscopy common problems a mixtures

Example 6: The equilibrium constant for the conjugate acid-base pair

HIn + H2O H3O+ + In-

Calculate the absorbance at 430 nm for an indicator concentration of 3.00x10-4 M

K = 8.00x10-5

e = 0.755x103

e = 8.04x103


Application of uv vis spectroscopy common problems a mixtures

Applications:

B) Absorbing Species in UV/Vis:

1)Electronic transitions involving organic compounds, inorganic compounds, complexes, etc.

Basic process:

M + hn M*

10-8 – 10-9s

M*  M + heat

(or fluorescence, light, or phosphorescence)

or

10-8 – 10-9s

M*  N

(new species, photochemical reaction)

Note: excited state (M*) is generally short and heat produced not generally measurable.

Thus, get minimal disturbance of systems (assuming no photochemical reaction)


Application of uv vis spectroscopy common problems a mixtures

2)Absorption occurs with bonding electrons.

- E(l) required differs with type of bonding electron.

- UV-Vis absorption gives some information on bonding electrons (functional groups in a compound.

- Most organic spectra are complex

‚electronic and vibration transitions superimposed

‚absorption bands usually broad

‚detailed theoretical analysis not possible, but semi-quantitative or qualitative analysis of types of bonds is possible.

‚ effects of solvent & molecular details complicate comparison


Application of uv vis spectroscopy common problems a mixtures

- Single bonds usually too high excitation energy for most instruments (£185 nm)

‚vacuum UV

‚most compounds of atmosphere absorb in this range, so difficult to work with.

‚usually concerned with functional groups with relatively low excitation energies (190 £l £850 nm).

- Types of electron transitions:

i) s, p, n electrons

Sigma (s) – single bond electron

Low energy bonding orbital

High energy anti-bonding orbital


Application of uv vis spectroscopy common problems a mixtures

Pi (p) – double bond electron

Low energy bonding orbital

High energy anti-bonding orbital

Non-bonding electrons (n): don’t take part in any bonds,

neutral energy level.

Example:Formaldehyde


Application of uv vis spectroscopy common problems a mixtures

‚s s* transition in vacuum UV

‚n s* saturated compounds with non-bonding electrons

l ~ 150-250 nm

e ~ 100-3000 ( not strong)

‚n p*, p  p* requires unsaturated functional groups (eq. double bonds)

most commonly used, energy good range for UV/Vis

l ~ 200 - 700 nm

n p* : e ~ 10-100

p  p*: e ~ 1000 – 10,000


Application of uv vis spectroscopy common problems a mixtures

Absorption Characteristics of Some Common Chromophores


Application of uv vis spectroscopy common problems a mixtures

Other Examples of Some Common Chromophores


Application of uv vis spectroscopy common problems a mixtures

ii) d/f electrons (transition metal ions)

‚ Lanthanide and actinide series

- electronic transition of 4f & 5f electrons

- generally sharp, well-defined bands not affected by associated ligands

‚ 1st and 2nd transition metal series

- electronic transition of 3d & 4d electrons

- broad peaks

Crystal-Field Theory

  • In absence of external field d-orbitals are identical

  • Energies of d-orbitals in solution are not identical

  • Absorption involves e- transition between d-

    orbitals

  • In complex, all orbitals increase in energy where

    orbitals along bonding axis are destabilized


Application of uv vis spectroscopy common problems a mixtures

Magnitude of D depends on:

- charge on metal ion

- position in periodic table

- ligand field strength :

I- < Br- < Cl- < F- < OH- < C2O42- ~ H2O < SCN- < NH3

< ethylenediamine < o-phenanthroline < NO2- < CN-

D increases with increasing field strength, so wavelength decreases


Application of uv vis spectroscopy common problems a mixtures

iii) Charge Transfer Complexes

‚ Important analytically because of large e (> 10,000)

‚Absorption of radiation involves transfer of e- from the donor to orbital associated with acceptor

- excited state is product of pseudo oxidation/reduction process

‚ Many inorganic complexes of electron donor (usually organic)

& electron acceptor (usually metal)

- examples: Iron III thiocyanate

Iron II phenanthroline

(colorless)

(deep red color)


Application of uv vis spectroscopy common problems a mixtures

C) Qualitative Analysis:

1)Limited since few resolved peaks

- unambiguous identification not usually possible.

2) Solvent can affect position and shape of curve.

- polar solvents broaden out peaks, eliminates fine structure.

Loss of fine structure for acetaldehyde when transfer to solvent from gas phase

Also need to consider absorbance of solvent.


Application of uv vis spectroscopy common problems a mixtures

2) Solvent can affect position and shape of curve.

- polar solvents broaden out peaks, eliminates fine structure.

(a) Vapor

Loss of fine structure for 1,2,4,5-tetrazine as solvent polarity increases

(b) Hexane solution

(c) Aqueous


Application of uv vis spectroscopy common problems a mixtures

3) Solvent can also absorb in UV-vis spectrum.


Application of uv vis spectroscopy common problems a mixtures

3) Can obtain some functional group information for certain types of compounds..

- weak band at 280-290 nm that is shifted to shorter l’s with an increase in polarity (solvent) implies a carbonyl group.

‚acetone:

in hexane, max = 279 nm ( = 15)

in water, max = 264.5 nm

- solvent effects due to stabilization or destabilization of ground or excited states, changing the energy gap.

‚since most transitions result in an excited state that is more polar than the ground state

- 260 nm with some fine structure implies an aromatic ring.

Benzene in heptane

More complex ring systems shift to higher l’s (red shift) similar to conjugated alkenes


Application of uv vis spectroscopy common problems a mixtures

UV Spectral Nomenclature


Application of uv vis spectroscopy common problems a mixtures

C) Quantitative Analysis (Beer’s Law):

1)Widely used for Quantitative Analysis Characterization

- wide range of applications (organic & inorganic)

- limit of detection  10-4 to 10-5 M (10-6 to 10-7M; current)

- moderate to high selectivity

- typical accuracy of 1-3% ( can be ~0.1%)

- easy to perform, cheap

2) Strategies

a) absorbing species

- detect both organic and inorganic compounds containing any of these species (all the previous examples)


Application of uv vis spectroscopy common problems a mixtures

b) non- absorbing species

- react with reagent that forms colored product

- can also use for absorbing species to lower limit of detection

- items to consider:

l, pH, temperature, ionic strength

- prepare standard curve (match standards and samples as much as possible)

Non-absorbing

Species (colorless)

reagent

(colorless)

Complex

(red)

When all the protein is bound to Fe3+,

no further increase in absorbance.

As Fe3+ continues to bind protein

red color and absorbance increases.


Application of uv vis spectroscopy common problems a mixtures

Standard Addition Method (spiking the sample)

- used for analytes in a complex matrix where interferences in the UV/Vis for the analyte will occur: i.e. blood, sediment, human serum, etc..

- Method:

(1) Prepare several identical aliquots, Vx, of the unknown sample.

(2) Add a variable volume, Vs, of a standard solution of known concentration, cs, to each unknown aliquot.

Note: This method assumes a linear relationship between instrument response and sample concentration.


Application of uv vis spectroscopy common problems a mixtures

(3) Dilute each solution to an equal volume, Vt.

(4) Make instrumental measurements of each sample to get an instrument response, IR.


Application of uv vis spectroscopy common problems a mixtures

(5) Calculate unknown concentration, cx, from the following equation.

Note: assumes a linear relationship between instrument response and sample concentration.

Where:

S = signal or instrument response

k =proportionality constant

Vs =volume of standard added

cs =concentration of the standard

Vx =volume of the sample aliquot

cx =concentration of the sample

Vt =total volume of diluted solutions


Application of uv vis spectroscopy common problems a mixtures

c) Analysis of Mixtures

- use two different l’s with different e’s

A1 = e1MbcM + e1NbcN(l1)

A2 = e2MbcM + e2NbcN(l2)

Note: need to solve simultaneous equations


Application of uv vis spectroscopy common problems a mixtures

d) Photometric titration

-can measure titration with UV-vis spectroscopy.

- requires the analyte (A), titrant (T) or titration product (P) absorbs radiation


Application of uv vis spectroscopy common problems a mixtures

Example 7: Given:

Calculate the concentrations of A and B in solutions that yielded an absorbance of 0.439 at 475 nm and 1.025 at 700 nm in a 2.50-cm cell.