RHIC pC Setup Run11

Loading in 2 Seconds...

play fullscreen
1 / 9

RHIC pC Setup Run11 - PowerPoint PPT Presentation


  • 98 Views
  • Uploaded on

RHIC pC Setup Run11. Polarimeter Status. 111 bunch injection studies . Work during access on Wednesday: fixed problem with readout of yellow-1-downstream polarimeter investigated problem with current monitoring of Y2U spares of modules involved located

loader
I am the owner, or an agent authorized to act on behalf of the owner, of the copyrighted work described.
capcha
Download Presentation

PowerPoint Slideshow about ' RHIC pC Setup Run11' - caroun


An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -
Presentation Transcript
rhic pc setup run11
RHIC pC Setup Run11

RSC-Meeting, January 21st 2011

polarimeter status
Polarimeter Status

111 bunch injection

studies

  • Work during access on Wednesday:
    • fixed problem with readout of yellow-1-downstream polarimeter
    • investigated problem with current monitoring of Y2U
      • spares of modules involved located
      • can hopefully fix the problem at the next access day

Hardware status okay

  • Change in operational procedure:
    • turn off Si bias voltage during beam studies, tune-up, APEX
      • keep radiation damage at a minimum

RSC-Meeting, January 21st 2011

first data from b1 b2 polarimeters
First Data from B1 & B2 Polarimeters

Blue-1 Upstream

  • Raw Data:
  • bananas clearly visible
  • calibration ADC -> energy
    • alpha calibrations
  • need to correct for Si – dead layer

Blue-2 Downstream

RSC-Meeting, January 21st 2011

a calibration
a-Calibration

Blue-1 Upstream

factor 2 in position comes from the

fact that downstream polarimeters

don’t use shapers,

which also amplify

Remember

a’s penetrate much deeper into Si

than carbons

  • Plan for upcoming shutdown:
  • install 2nd source with different
  • - energy
  • install different foils to attenuate
  • a – energy (tests in lab planned)

Blue-2 Downstream

RSC-Meeting, January 21st 2011

data after corrections
Data after Corrections

t0

Blue-1 Upstream

Blue-1 Upstream

energy-scale correction

based on Si dead layer

Energy-Scale

Blue-2 Downstream

Blue-2 Downstream

Current Puzzle:

difference between t0 and

energy scale correction factor

RSC-Meeting, January 21st 2011

data after all corrections
Data after all Corrections

Blue-1 Upstream

After all corrections both

polarimeters look very similar

Are you ready ????

Fasten your seat belts

the secret will be revealed

Blue-2 Downstream

RSC-Meeting, January 21st 2011

polarization at injection for blue ring
Polarization at Injection for Blue Ring

14807.003: blue1 upstream P = 51 +/- 3 % (stat only)

14807.212: blue2 downstream P = 49 +/- 3 % (stat only)

RSC-Meeting, January 21st 2011

to do
To Do
  • Update parameters in online polarization calculation to get reasonable numbers
  • Investigate the difference in energy-scale constant between B1 and B2
    • see we get the same result for Y1 and Y2 (just collected the data last night)
    • measure linearity of energy response of different systems
      • vary pulser amplitude  latest during the next access day
  • Need good data for Y1 and Y2 polarimeters to determine parameters for online, most likely the ones from last night will do.
  • Analyze the polarization at flat top for B2
  • Next access-day
    • fix bias voltage readout
    • bring hopefully one detector back to live in Y2U by changing pre-amp box

RSC-Meeting, January 21st 2011