Lisst sl plots
Download
1 / 24

LISST-SL plots - PowerPoint PPT Presentation


  • 101 Views
  • Uploaded on

LISST-SL plots. R0681302. Pump takes ~2 mins to lock in Data no good. R1091106. No apparent issue with pump voltage. Apart from the 2 high conc values and open-ended PSD data looks good. R1101437. Pump not locked on for the first minute; possibly lock-on after 1 min?

loader
I am the owner, or an agent authorized to act on behalf of the owner, of the copyrighted work described.
capcha
Download Presentation

PowerPoint Slideshow about ' LISST-SL plots' - bryga


An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -
Presentation Transcript

R0681302
R0681302

Pump takes ~2 mins to lock in

Data no good


R1091106
R1091106

No apparent issue with pump voltage.

Apart from the 2 high conc values and open-ended PSD data looks good


R1101437
R1101437

Pump not locked on for the first minute; possibly lock-on after 1 min?

Upcast probably OK. Note constant tau, but variation in MC. PSD looks fine.


R1101500
R1101500

Pump not locked on. Data not OK.


R1120717
R1120717

Velocity very low, so pump shuts off then starts up again.


R1120727
R1120727

Pitot tube contamination, not reliable data.


R1121523
R1121523

Pump resetting. Data not OK.


R1121618
R1121618

Pump resetting after ~1 minute. First minute may be OK..


R1160408
R1160408

Pump seems to track towards the end, last ~10 data points should be OK.


R1301058
R1301058

Pump never locked in. Data no good (velocity, depth OK).

Why max velocity in the middle?


R1301440
R1301440

Pump never locked in. Data no good (velocity, depth OK).


R1301447
R1301447

Excellent pump lock-in and tracking of velocity. Good PSD, good concentration.


R1311450
R1311450

Excellent pump lock-in and tracking of velocity.


R1311507
R1311507

Good pump lock-in for the last half.




Senachwine creek at chillicothe
Senachwine Creek at Chillicothe


Vc vs mc linear
VC vs MC, linear


Vc vs mc log
VC vs MC, log


M c vc vs tau
MC/VC vs tau


M c vc vs velocity
MC/VC vs velocity


M c vc vs mass 0 0625mm
MC/VC vs Mass%<0.0625mm


M c vc vs volume 0 0625mm
MC/VC vs Volume%<0.0625mm


ad