1 / 6

Acuson/Siemens V5Ms Cardiac Phased array probe image degradation

Probelogic Exclusive : We repair medical ultrasound transducer. Contact us for probe care, repair, testing, replacements.<br>It is a Siemens/Acuson V5Ms TOE/TEE probe in which the customer was complaining about poor image quality. Our acoustic report suggests the probe has approximately 10% of elements non firing as can be seen in the image below.<br>The issue is almost certainly caused by a significant knock to the head of the probe as can be seen in the image below.<br>Now if we compare a good V5Ms ultrasound image from the same phantom on the same ultrasound system. Firstly the good image, you can see how clear the phantom dots throughout the image are:<br>With such gradual deterioration, it is sometimes difficult to recognize the problem as it just gets more and more difficult to scan patients. Get your probes acoustically tested today to be sure.<br> <br>Learn more on Acuson/Siemens V5Ms Cardiac Phased array probe image degradation<br>http://probelogic.com.au/acusonsiemens-v5ms-cardiac-phased-array-probe-image-degradation/<br> <br>Contact Probelogic today<br>Opening Hours : Monday to Friday - 8:30am to 5:00pm<br>Contact : 1300-611-503

bryanlen
Download Presentation

Acuson/Siemens V5Ms Cardiac Phased array probe image degradation

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Acuson/Siemens V5Ms Cardiac Phased array probe image degradation Subscribe to probelogic blog

  2. Acuson/Siemens V5Ms Cardiac Phased array probe image degradation Subscribe to probelogic blog Currently we have a probe in our laboratory for investigation. It is a Siemens/AcusonV5Ms TOE/TEE probe in which the customer was complaining about poor image quality. Our acoustic report suggests the probe has approximately 10% of elements non firing as can be seen in the image.

  3. Acuson/Siemens V5Ms Cardiac Phased array probe image degradation Subscribe to probelogic blog The issue is almost certainly caused by a significant knock to the head of the probe as can be seen in the image.

  4. Acuson/Siemens V5Ms Cardiac Phased array probe image degradation Subscribe to probelogic blog Now if we compare a good V5Ms ultrasound image from the same phantom on the same ultrasound system. Firstly the good image, you can see how clear the phantom dots throughout the image are:

  5. Acuson/Siemens V5Ms Cardiac Phased array probe image degradation Subscribe to probelogic blog Now comparing with an image taken from the faulty probe: With such gradual deterioration, it is sometimes difficult to recognize the problem as it just gets more and more difficult to scan patients. Get your probes acoustically tested today to be sure.

  6. Acuson/Siemens V5Ms Cardiac Phased array probe image degradation Subscribe to probelogic blog http://probelogic.com.au/acusonsiemens-v5ms-cardiac-phased-array-probe-image-degradation/

More Related