Gain measurement with LED and time-over-threshold

Gain measurement with LED and time-over-threshold. ADC. passive splitter. Analog out. Threshold L. Threshold H. TDC. LED. LED driver. LVDS out.

berny
Download Presentation

Gain measurement with LED and time-over-threshold

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Gain measurement with LED and time-over-threshold ADC passive splitter Analog out Threshold L Threshold H TDC LED LED driver LVDS out

  2. It is possible to span different regions of signal amplitude by modulating the charge (voltage and time-width) to the LED (with the CAMAC LED driver) A small variation in the charge injected into the LED produces a large difference in the produced light (and thus in the signal amplitude) The intrinsic fluctuation due to the LED alone is very small, compared to the s of the signal due to the gain of the PMT

  3. Original divider

  4. Modified divider • Reconstruct ToT vs. Q calibration curve using Q from ADC vs. ToT from TDC

  5. Compare two different thresholds: - As expected, higher threshold results in steeper curve

  6. Charge from ADC Charge from TDC Charge from ADC given a TDC hit

  7. Fit gain using Q from ADC • Using ToT vs. Q calibration curve reconstruct Q from ToT • Fit gain with usual method

  8. We have seen that is possible to use ToT vs. Q calibration curve to reproduce the gain measurement, for a single crystal • Now we want to: • Produce calibration curve for 1 crystal • Use this curve for several other crystals • Try to reproduce gain with ToT • Setup ready for 6 lead-glasses

  9. Small problem: find LED’s with similar luminosity • (in order to span similar charge/ToT regions)