HAPPEX/ Lumi PMT Studies. Luis Mercado UMASS 7/10/09. PMT Linearity. Past tests have shown nonlinearities in PMT output dependant on input and gain. Want to study and document good linearity regions for Lumi and HAPPEX PMTs. PMT Linearity (Formalism).

Download Presentation

HAPPEX/ Lumi PMT Studies

An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -

Presentation Transcript


Happex lumi pmt studies

HAPPEX/Lumi PMT Studies

Luis Mercado

UMASS

7/10/09


Pmt linearity

PMT Linearity

  • Past tests have shown nonlinearities in PMT output dependant on input and gain.

  • Want to study and document good linearity regions for Lumi and HAPPEX PMTs.


Pmt linearity formalism

PMT Linearity (Formalism)

  • A PMT’s non-linear response can be expressed as follows:

  • Experimental asymmetry can be approximated:


Experimental test setup

Experimental Test Setup


Lumi pmt linearity

Lumi PMT Linearity

Input = 12nA

HV = 370

Output = uA

B*N_0 = -5.14%


Good lumi pmt linearity

Good Lumi PMT Linearity

Input =~1nA

HV = 690

Output ~ 7uA

B*N_0 = 1.19%


Happex pmts

HAPPEX PMTs

  • Received one HAPPEX PMT last week.

  • Re-built experimental setup with bigger box.

  • Have done linearity tests at input photocathode currents of [0.06-6.0] nA.

  • Only good linearity region found was for an input of ~1nA at HV=1250.


Happex lumi pmt studies

  • Linearity of output signal breaks down with high enough HV.

  • Signal becomes unpredictable in these regions.


Happex pmt linearity

HAPPEX PMT Linearity

Input ~1nA

HV = 1100

Output = 7uA

B*N_0 = -15.69%


Good happex pmt linearity

Good HAPPEX PMT Linearity

Input ~ 1nA

HV = 1250

Output~ 37 uA

B*N_0 = 0.35%


Odd pedestal behavior

Odd Pedestal Behavior

  • The pedestal signal is huge for HV=1250.

  • Huge signal with no light.


Odd pedestal behavior1

Odd Pedestal Behavior

  • Also happens with Lumi PMT.


Ongoing tests

Ongoing Tests

  • Need to track down the cause of the odd pedestal behavior.

  • Study other available HAPPEX PMT.

  • Finalize linearity mapping for all PMTs.


Acknowledgements

Acknowledgements

  • Krishna Kumar

  • Kent Pashke

  • Dustin McNulty

  • Robert Michaels

  • Aimee Shore